Page 97
Appendix |ASCII Communication with Controller
confocalDT 24x1
A 5.3.2.5 Changing the Password
PASSWD <Old Password> <New Password> <New Password>
Changes the password for the PROFESSIONAL level. The default (preset) password is
“000”.
The old password must be entered once, and the new password twice. If the new pass-
words do not match, an error message is displayed. Passwords are case sensitive. A
password may only contain letters (A to Z) and numbers, but no special characters and
no letters with accents or umlauts. The maximum length is 31 characters.
A 5.3.3 Sensor
A 5.3.3.1 Info about Calibration Tables
SENSORTABLE
->SENSORTABLE
Pos, Sensor name,
Range,
Serial
0, ifs-2405x,
3,000mm,
12345678
8, ifs-2405x,
10,000mm,
12345678
9, ifs-2405x,
3,000mm,
12345678
->
All available (learned) sensors are displayed.
A 5.3.3.2 Sensor Number
SENSORHEAD <sensor position>
Selects the current sensor from its position, see Chapter
Minimum 0, maximum 19.
Error
Description
E39
Sensor is not available
A 5.3.3.3 Sensor Information
SENSORINFO
Displays sensor data (name, measuring range and serial number).
->SENSORINFO
Position:
0
Name:
ifs-2405x
Measuring range:
3.000 mm
Serial:
12345678
->
A 5.3.3.4 Dark Correction
DARKCORR
Performs a dark correction. Dark correction is sensor-dependent and is stored sepa-
rately in the controller for each sensor. Before carrying out a dark correction, select the
required sensor.
A 5.3.3.5 Warning Threshold in the Event of Contamination
DARKCORRTHRES <threshold>
Threshold: Deviation (in %) of the intensity/exposure time from the stored value, above
which a warning message will appear. Default setting: 50 %.
The warning threshold is set in % with one decimal.