Page 106
Appendix |ASCII Communication with Controller
confocalDT 24x1
A 5.5.2 Select Measurement Values to be Output
Setting the values to be output via the RS422 and Ethernet interface.
Maximum 32 measurement values are transmitted with RS422 in parallel.
The maximum output rate via the Ethernet interface depends on the number of output
values.
Any distances and differences can be selected for the output in the multilayer measure-
ment mode. All measurement values required for the difference calculations are output in
addition to the Ethernet measuring value transmission.
Via the Ethernet interface always the displacement 1 and in case of thickness measure-
ment the displacement 1 and 2 and the difference 1-2 is output .
A 5.5.2.1 Data Selection Displacement Measurement for RS422 / Ethernet
Controller IFC24x1
Controller IFC24x1MP
OUTDIST_RS422 NONE|[DIST1]
[DIST2]
OUTDIST_ETH NONE|[DIST1]
[DIST2]
OUTDIST_RS422 NONE|[DIST1]
[DIST2] [DIST3] [DIST4]
[DIST5] [DIST6]
OUTDIST_ETH NONE|[DIST1]
[DIST2] [DIST3] [DIST4]
[DIST5] [DIST6]
Setting, which displacement values are output through RS422 or Ethernet. All active
distance and thickness values are transmitted via Ethernet.
-
NONE: No output of a displacement
-
DIST1: Output of displacement 1
-
DIST2: Output of displacement 2
In measurement mode
Thickness measurement
the displacement 1 and/or 2 can be
output.
The selection of the distances 1 or 2 can be done only in the video mode when output
via Ethernet. The other modes transmit the values according to the selected measure-
ment program.
In
distance
measurement mode the controller accepts the
OUTDIST_ETH DIST1
command (no output with
NONE
).
In
thickness
measurement mode the controller accepts the
OUTDIST_ETH DIST1
DIST2
command (no output with
NONE
, it must always be both).
In
multi peak
measurement mode the controller accepts the
OUTDIST_ETH
DIST1|...
command (no output with
NONE
, it must always be all distances required for
thickness calculation).
A 5.5.2.2 Data Selection Thickness Measurement for RS422 / Ethernet
Controller IFC24x1
Controller IFC24x1MP
OUTTHICK_RS422 NONE|[THICK12]
OUTTHICK_ETH NONE|[THICK12]
OUTTHICK_RS422 NONE|[THICK12]
[THICK13][THICK14] [THICK15]
[THICK16][THICK23][THICK24]
[THICK25][THICK26][THICK34]
[THICK35][THICK36][THICK45]
[THICK46][THICK56]
OUTTHICK_ETH NONE|[THICK12]
[THICK13][THICK14] [THICK15]
[THICK16][THICK23][THICK24]
[THICK25][THICK26][THICK34]
[THICK35][THICK36][THICK45]
[THICK46][THICK56]
Defines, which calculated layer thicknesses are output via RS422 or Ethernet.
-
NONE: No output of calculated layer thickness
-
THICK12: Output of the layer thickness between displacement 1 and 2