
Page 87
Appendix | Control Menu
C-Box/2A
A 3.2.2
Data Recording
Measurement
task
Measuring
mode
Measurement value
sensor 1
Measured value of the sensor connected to connection 1, i.e., the C-Box/2A
value includes the value from sensor 1
Thickness sensor
1-2
Forms the difference between the two distance values of the sensors 1/2 in
direct or diffuse reflection, with two-sided distance measurement, and outputs
the result as a thickness value
Step sensor 1-2
Forms the difference between the two distance values of the sensors 1/2 in
direct or diffuse reflection, with one-sided distance measurement, and outputs
the result as height value
Measuring
rate
Measuring rate
(kHz)
0.5 / 1.0 / 2.0 / 4.0
When synchronization is switched off, the measuring rate can be set freely.
Value range: from 0.4 to 80 kHz. Otherwise, the possible measuring rates are
specified by the connected sensors / controllers
Data rate Web-
diagram (kHz)
0.5 / 1.0
i
The Web diagram interface uses a slower data rate for data transmis-
sion. So for higher measuring rates not all the measured values will be
visible in the diagram or saved to file.
Error handling
Error handling on
wrong measure-
ment values
Error output, no value
If a valid metric can not be obtained, an error value is output.
If this hinders further processing, alternatively the last valid
measured value can be held over a certain number of measur-
ing cycles, that is, repeatedly output
Hold last valid value
Value
Hold last valid value forever
Содержание C-Box/2A
Страница 95: ......