Appendix | ASCII Communication with Sensor
C-Box/2A
A 2.3.3
Interfaces
A 2.3.3.1 Ethernet
IPCONFIG DHCP|STATIC [<IPAdresse> [<Netmask> [<Gateway>]]]
Set Ethernet interface.
- DHCP: IP address and gateway are automatically requested by DHCP. System looks for a LinkLocal ad-
dress after appr. 30 minutes if no DHCP server is available.
- STATIC: Set IP address, net mask and gateway in format xxx.xxx.xxx.xxx
Values stay the same if no IP address, net mask, and/or gateway is typed in.
Default = STATIC 169.254.168.150 255.255.0.0 169.254.1.1
A 2.3.3.2 Setting the Measured Value Server
MEASTRANSFER SERVER/TCP [<PORT>]
I
n case of measured value output via Ethernet: currently only TCP server is provided.
- The port is freely selectable between 1024 and 65535.
Default = SERVER/TCP 1024
A 2.3.3.3 Baudrate
BAUDRATE <Baudrate>
Setting the interface baudrate to the PC. Possible variants: 115.200 (Default), 8.000.000, 4.000.000,
3.500.000, 3.000.000, 2.500.000, 2.000.000, 1.500.000, 921.600, 691.200, 460.800, 230.400, 9.600 Baud
Default = 115200
A 2.3.3.4 Find C-Box/2A
Search the C-Box/2A by using the
sensorTOOL
x.x.x
program
A 2.3.4
Handling of Setups
A 2.3.4.1 Save Parameter
STORE 1|2|3|4|5|6|7|8
Save the current parameter under the specified number in the flash. With the restart of the C-Box/2A the last
saved data record is always loaded.