Anhang | ASCII Communication with Sensor
optoNCDT ILR2250
A 3.5.5 ERRORLIMITVALUES1/2/3
ERRORLIMITVALUES1 [<lower> <upper>]
ERRORLIMITVALUES2 [<lower> <upper>]
ERRORLIMITVALUES3 [<lower> <upper>]
Sets the lower and upper limit value for the switching outputs.
Value range:
<lower> Value range between 0.0 and 150,000.0 (mm, one decimal place)
<lower> Value range between 0.0 and 150,000.0 (mm, one decimal place)
A 3.5.6 ERRORHYSTERESIS1/2/3
ERRORHYSTERESIS1 [<hysteresis>]
ERRORHYSTERESIS2 [<hysteresis>]
ERRORHYSTERESIS3 [<hysteresis>]
Value by which the measured value must be above or below the limit in order for the switching output to be triggered or
deactivated again.
<hysteresis>: Value range between 0.0 and 150,000.0 (mm, one decimal place)
A 3.5.7 ERROROUTHOLD
ERROROUTHOLD [<time>]
Specification of the minimum time in ms for which the switching output should remain active if the limit value is exceed-
ed. The time period begins when the limit value is exceeded.
<time> Value range between 0 and 10000 ms.
A 3.5.8 OUTHOLD, Error Handling
OUTHOLD [NONE|0|<count>]
Sets the measured value output behavior in the event of an error.
-
NONE: Last measured value not held; error value output.
-
0: Last measured value held indefinitely
-
<count> Last measured value held for n measurement cycles, afterwards an error value is output.
n = (1 ... 2147483645).
A 3.6
Setup and Measurement Settings
A 3.6.1 MEASSETTINGS
MEASSETTINGS [READ|STORE|PRESETLIST|PRESETMODE [<mode>]]
Settings for measurement task. Loads manufacturer-defined presets or user-specific setups from the sensor or saves
user-specific setups in the sensor.
-
READ: Read measurement settings from non-volatile memory.
-
STORE: Write measurement settings to non-volatile memory.
-
PRESETLIST: Lists all existing configurations
-
PRESETMODE: Sets a configuration
-
<mode>: Name of a configuration listed with the subcommand
PRESETMODE
.
A 3.6.2 BASICSETTINGS
BASICSETTINGS [READ|STORE]
-
READ: Loads the saved device settings from the sensor.
-
STORE: Saves the current device settings in the sensor.