Page 29
Installation and Initial Operation
thicknessCONTROL MTS 8201.LLT / 4350006.71
Signal name
-7X1
contact
Customer
cable
emergency stop
contact : 2
3
emergency stop
contact : 2
4
PE
5
Connect the cable for the Analog signal to +ET-15A3:
Signal name
+ET-15A3
contact
Customer
cable
Analog signal
-10V…10V
O1
GND
-
Shield
Shielding rail cabinet
Connect Ethernet IP to +ET1-9X5.
Connect remote support (VPN) to +ET1-9X6.
Connect EtherCAT cable -4W1 to +ET1-4E1:X4 and to +OFR-4X1.
Connect Sensor 1 Signal cable -9W4 to +ET1-9X1 and to +OFR-9X2.
Connect Sensor 2 Signal cable -9W7 to +ET1-9X3 and to +OFR-9X4.
Connect 24V/Dig I/O cable -28W1 to +ET1-28X1 and to +OFR-28X2.
Connect Resolvercable -4W2.1 to +ET1-4E1:X7 and to +OFR-4X2.
Connect Motorcable -4W3.1 to +ET1-4E1:X107 and to +OFR-4X3.
Connect the grounding terminal of the measuring system to the potential equalization of the electrical installation
using a potential equalization conductor of at least 10 mm².
4.6
Compressed Air Supply
The connection for the compressed air supply is located on the measuring system.
Connect the compressed air supply
The following properties apply:
Input pressure
Nominal pressure max. 6 bar
±5 %
, dry
and oil-free
Air consumption
30 l/min
Cleanliness
ISO8573-1:2010 [1:4:2]
4.7
Switching on and off Process
4.7.1
Switching on Electrical Components
The system is switched on using the main switch. The PC starts the boot process. This is completed with the start of
the measuring program. The machine operator must login to the system. The system is now ready to use.
4.7.2
Switching off Electrical Components
The measuring program must be ended in order to shut down the system properly. The PC is shut down automatical-
ly and switched off. The main switch must be switched off to disconnect the system from the power supply.
Содержание 4350006.71
Страница 1: ...thicknessCONTROL MTS 8201 LLT O Frame Non contact thickness measurement Operating Instructions ...
Страница 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 95: ...Operating Instructions scanCONTROL 29xx ...
Страница 99: ...scanCONTROL 29xx Appendix A 1 Accessories 60 A 1 1 Recommended Accessories 60 A 1 2 Optional Accessories 61 ...
Страница 100: ......
Страница 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Страница 193: ......
Страница 194: ......
Страница 229: ......