Operating Manual 

 

Federal (USA) law restricts this device to sale by or on 

the order of a physician or licensed practitioner. 

 

 

Micro Direct, Inc. 

803 Webster Street 

Lewiston, ME 04240 

1-800-588-3381 

www.mdspiro.com 

 
 

MAN1281 

Issue 1.1 

December 2012 

Содержание SmokeCheck

Страница 1: ...al Federal USA law restricts this device to sale by or on the order of a physician or licensed practitioner Micro Direct Inc 803 Webster Street Lewiston ME 04240 1 800 588 3381 www mdspiro com MAN1281...

Страница 2: ...Operation Page 6 Calibration Page 9 Battery Life Page 11 Battery Replacement Page 12 Sensor Expiry Page 13 Cleaning Page 13 Servicing Page 14 Trouble Shooting Information Page 15 Safety Designation pe...

Страница 3: ...d personnel Introduction The Smoke Check is based on an electrochemical fuel cell sensor which works through the reaction of carbon monoxide CO with an electrolyte at one electrode and oxygen from amb...

Страница 4: ...ograms Measurement of expired CO has been well validated as an indirect measure of cigarette consumption and is widely used in smoking cessation programs Typical values for expired CO in smokers toget...

Страница 5: ...by CO measurement Respiratory and TB Unit General Hospital Via forlanini 71 I 31029 Vittorio Veneto TV ITALY Marino Luigi Latini Roberto Barbano Gina Bazzerla Zanette Antonia Nardini Stefano Correlati...

Страница 6: ...4...

Страница 7: ...e batteries connect improperly or dispose of in a fire as there is a possibility of leakage or explosion Follow manufacturer s recommendations for proper disposal WARNING The instrument is not suitabl...

Страница 8: ...rior to use Install the 9 volt battery by sliding open the battery cover clipping the battery in place and replacing the cover Insert the mouthpiece adapter into the Smoke Check meter and then insert...

Страница 9: ...ubject to inspire and hold their breath The unit will auto zero for twelve seconds during which time the following will be displayed W a it Important note the unit must not be used during this period...

Страница 10: ...indicator 11 20 ppm with red indicator 20 ppm with flashing red indicator and alarm Important note Before repeating a measurement the unit must be turned off and the mouthpiece and adapter removed fo...

Страница 11: ...this indicates possible contamination of the sensor with a solvent In this case remove all sources of solvent from the vicinity of the sensor and expose to ambient air for 24 hours before switching o...

Страница 12: ...ow I nd i ca t o r Con tr o l Va l ve P l as ti cTub i ng Locate the calibration button positioned inside a hole on the right hand side of the instrument as show below Turn the unit on and wait for t...

Страница 13: ...all the legends and the three lights will flash five times with no beeps after the calibration button is depressed The GAS legend will be displayed as before and the unit must be turned off before rec...

Страница 14: ...r situated on the rear of the unit towards the bottom of the device Place your thumb over the round thumb indent press gently and slide the cover to the right to remove it from the unit Lift the old b...

Страница 15: ...nts such as alcohol and cleaning fluids When the sensor has expired it will become impossible to obtain a correct calibration When this occurs the Smoke Check must be returned to Micro Direct for sens...

Страница 16: ...ohol or permanent damage may result CAUTION Do not attempt to wash or immerse the Smoke Check meter in water or cleaning fluid as there are electronic components inside that will be permanently damage...

Страница 17: ...ry life is shorter than expected Unit not being switched off Turn the unit off after use GAS is displayed Fuel cell is depleted Return unit for cell replacement Fuel cell is contaminated by residual g...

Страница 18: ...lassifications according to IEC 60601 1 Smoke Check Applied part type B WARNING No modification of this equipment is allowed WARNING Do not connect devices that are not specified as part of the system...

Страница 19: ...oke Check monitor is configured and installed put into service in accordance with the instructions guidance provided herein and is used only in the configuration as supplied Changes or modifications t...

Страница 20: ...nded for use in the electromagnetic environment specified below It is the responsibility of the customer or the user of the Smoke Check to ensure that is used in such an environment Immunity Test IEC...

Страница 21: ...ce and Manufacturer s Declaration Electromagnetic Immunity The Smoke Check is intended for use in the electromagnetic environment specified below It is the responsibility of the customer or the user o...

Страница 22: ...ion in which the Smoke Check is used exceeds the applicable RF compliance level above the Smoke Check should be observed to verify normal operation If abnormal performance is observed additional measu...

Страница 23: ...r manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affect...

Страница 24: ...e MC10 CO Calibration Gas 17 Liters MC15 CO Calibration Kit MC22 Regulator for Calibrating Gas MEC1184 Calibration Tool MEC1007 Reducing Connector for Calibration SSC5000 PeraSafe Sterilizing Powder F...

Страница 25: ...Sensor life 2 to 5 years Response time 15 sec to 90 of reading Hydrogen cross sensitivity 15 Operating temperature 0 104 O F Operating pressure Atmospheric 10 Pressure coefficient 0 02 signal per mBar...

Страница 26: ...chnical support inquiries please contact the following Micro Direct Inc 803 Webster Street Lewiston ME 04240 Customer Service and Sales Toll Free 800 588 3381 Telephone 207 786 7808 Email sales mdspir...