Micro Direct CO Check+ Скачать руководство пользователя страница 1

 

 

 

 
 
 

 

CO Check+  

Operating Manual 

 
 
 
 
 
 

Micro Direct, Inc. 

803 Webster Street 

Lewiston, ME 04240 

1-800-588-3381 

www.mdspiro.com

 

 
 

Revision 7.0 MD January 2015 

 

 
 
 
 
 

Содержание CO Check+

Страница 1: ...CO Check Operating Manual Micro Direct Inc 803 Webster Street Lewiston ME 04240 1 800 588 3381 www mdspiro com Revision 7 0 MD January 2015...

Страница 2: ...Warning Cautions 5 Calibration 6 Setting Environment Mode Optional 9 Battery Life 9 Power Saving 10 Cleaning 11 Servicing 11 Consumables 12 Specifications 13 Appendix A CO Check and Smoking Cessation...

Страница 3: ...pplied with 1 CO Check Cat No CO10 2 22 mm plastic mouthpiece adapter PSA1800 3 22 mm reducing connector for calibration MEC1007 4 9 Volt Battery 5 4 Disposable cardboard mouthpieces 3301 6 Soft shell...

Страница 4: ...ect s progress and compliance The CO Check is easy to use very accurate and requires a single breath into the device to display CO results in parts per million PPM and the percentage of carboxyhemoglo...

Страница 5: ...ently and continue blowing until their lungs are completely empty CO is collected in the last portion of the breath alveolar breath Test Reading Display The CO Check will beep and show the final readi...

Страница 6: ...o CO poisoning and medical attention should be immediately sort The CO Check has a cross sensitivity to hydrogen from some gastro intestinal disorders which may affect the reading Cleaning with produc...

Страница 7: ...6 Calibration The device should be recalibrated at least once every six months Calibration gas can be purchased from Micro Direct Inc see page 13 for full ordering details...

Страница 8: ...elect key 4 The device will countdown from 15 to 0 with CAL symbol ON It will then display the blow icon 5 Attach the connector and turn the control knob fully counter clockwise until knob will not tu...

Страница 9: ...dings within the range of 16 24 can be calibrated An Er Error message will be displayed if the calibration is accepted for reading outside this range 8 Switch off the device wait for a minute and it i...

Страница 10: ...At the end of the countdown the unit will display En for 3 seconds followed by either OF Off your current setting Press the Select key momentarily 0 5 sec to change the mode Press the Select key for...

Страница 11: ...e If the battery is very low such that the reading is not reliable it will display bd battery dead and will not operate any more Power Saving To save battery power the device will automatically turn i...

Страница 12: ...wipes see page 13 Please be certain not to touch the surface of the sensor or allow moisture to do the same Cleaning with products that contain alcohol may result in permanent contamination of the sen...

Страница 13: ...s MC22 Control Valve for Calibration 3301 Mouthpieces 100 per bag PSA1800 Intermediate plastic adaptors 10 per pack CS100 Replacement of fuel cell calibration battery and service GC01 CO Guide chart C...

Страница 14: ...Temperature 41 95 degrees F Operating Pressure Atmospheric 10 Operating Humidity 30 to 90 RH Operating Altitude Sea level to 6000 ft Storage Temperature 68 158 degrees F Storage Humidity 10 to 90 RH H...

Страница 15: ...3 42 EEC 0120 Type B Device Environment This instrument complies with directive EEC89 336 electromagnetic compatibility but may be affected by cellular phones and by electromagnetic interference excee...

Страница 16: ...er to see this CO monitoring is an extremely powerful tool for a smoking cessation counselor physician or respiratory therapist Taking a test before a COPD patient stops smoking will help you gauge th...

Страница 17: ...en when the person is taking a breath test this background reading of 10ppm will be taken into account For example if the Environmental monitor is ON and a reading of 10ppm is detected and the patient...