CO Check Pro
Operating Manual
Micro Direct, Inc
803 Webster Street
Lewiston, ME 04240
800-588-3381
[email protected]
www.mdspiro.com
Revision 10.1 January 2018
Страница 1: ...CO Check Pro Operating Manual Micro Direct Inc 803 Webster Street Lewiston ME 04240 800 588 3381 sales mdspiro com www mdspiro com Revision 10 1 January 2018...
Страница 2: ...rs 12 Battery Life 13 Power Saving 13 Cleaning 14 Servicing 14 Supplies 15 Specifications 15 Symbols 16 Important information regarding Electromagnetic Compatibility 17 Appendix A CO Check and Smoking...
Страница 3: ...stic mouthpiece adapter PSA1800 3 22 mm reducing connector for calibration MEC1007 4 9 Volt Battery 5 5 Disposable cardboard mouthpieces 3301 6 Hard shell plastic carry case CS101 7 Operating Manual K...
Страница 4: ...tion of carbon monoxide CO with an electrolyte at one electrode and oxygen from ambient air at the other This reaction generates an electrical current proportional to CO concentration Output from the...
Страница 5: ...key The device will display the version number before starting the countdown timer The breath hold timer will now count down from 15 seconds Encourage the subject to hold their breath until the devic...
Страница 6: ...battery cover and clipping the battery in place then replacing the battery cover Insert a SafeBreath mouthpiece order FM200 or if not using a SafeBreath mouthpiece insert the plastic mouthpiece adapte...
Страница 7: ...99 PPM Any reading above that will be displayed as which means it s over the range Colored lights are there to represent and provide a visual display of the resulting PPM reading The colored lights r...
Страница 8: ...faulty fireplace It does not meet the standards of protection of persons against CO pollution The SafeBreath filtered mouthpieces and regular disposable cardboard mouthpieces are single use only Re u...
Страница 9: ...hted option press the Select Function key for at least 3 seconds to confirm Calibration The device should be recalibrated at least once every six months Calibration gas can be purchased from Micro Dir...
Страница 10: ...s and hold the Select Function Key for at least 3 seconds to confirm the selection 5 The device will countdown from 15 to 0 and then will display the calibration icon 6 Attach the reducing connector t...
Страница 11: ...is now ready for use If the device is not calibrated for six months the device will beep three times before the countdown starts and will display the following message Calibration Expired Please Cali...
Страница 12: ...re e ea a at t th h h H H Ho o ol lld d d T T Ti iim m me e e Set Light colors Exit The first option is highlighted To select Breath Hold Time press the Function key until Breath Hold Time is highligh...
Страница 13: ...lo o or r rs s s Exit The first option is highlighted To select Set Light Colors press the Function key until Set Light Colors is highlighted To select press the Select Function key for at least 3 sec...
Страница 14: ...household waste Never dispose of batteries in fire because they could explode If the battery is so low that the reading is not reliable the device will display the battery dead icon and will not oper...
Страница 15: ...n then rinsed with water and left to dry thoroughly The device can be cleaned using non alcohol wipes see Supplies on page 15 Please be careful not to touch the surface of the sensor or allow any mois...
Страница 16: ...Detection Sensor Used Electrochemical Fuel Cell Sensitivity 0 1 PPM 0 10 PPM Range 1 0 PPM 10 99PPM Range Accuracy repeatability 2 ppm or 5 whichever is greater Operating Temperature 32 104O Fahrenhe...
Страница 17: ...dous substances in electrical and electronic equipment end users of electrical and electronic equipment should understand the meaning of the crossed out wheeled bin symbol Do not dispose of WEEE as un...
Страница 18: ...ck is intended for use in the electromagnetic environment specified below The customer or the user should assure that it is used in such an environment Immunity Test IEC 60601 Test level Compliance le...
Страница 19: ...O Check is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplied buildings used for domestic purposes RF...
Страница 20: ...ons Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless...
Страница 21: ...do a test is in the afternoon as CO levels fall overnight and morning readings can sometimes give misleadingly low results High readings in the morning on the other hand would be strong evidence of he...