14
3.2.2
Calibration
Run calibration wizard from
Calibration
menu of by using
tool button
Select calibration method –
Guided
(step-by-step manual procedure),
Automatic
(using
electronic calibrator module) or
Unguided
(fully manual calibration procedure) and press
Next
Select
Ports to Calibrate
(
Ports 1, 2
if transmission correction is needed), DUT connectors and
Calibration Kit
. Use
Calibration Kits…
button to import kit from a file to the list at the first run.
Supported connectors will be automatically added from the loaded kit. Or connect to electronic
calibrator in opened dialog.
Choose
Calibration Variant
(
SOLT
/
TRL
) and
Calibration Type
from the lists. Check desired
measurement options (
Isolation
,
Unknown THRU
, etc…) if available
Follow the wizard’s instructions to perform calibration
Click
Apply
to finish calibration procedure or
Back
to repeat one or more calibration steps
3.2.3
Measurements and Analysis
Connect the DUT
Adjust the scale manually or using
Autoscale
functions
Enable desirable trace functions
Analyze measured data with markers, their relations (which are created by connecting two
various markers with link, wired from mode icon ), limiting lines or statistics (Fig.3-3)
Prepare report using
Diagram
->
Report
menu
Save user settings by clicking
Profile
->
Save
menu item
Figure 3-3 Example of the bandpass filter measurement