80
☞
In doing so watch for cables that might still be at-
tached. – If required, remove corresponding cable
binders.
6. Remove the four lower screws of the lifting rod.
☞
While doing so have someone help you secure the
actuator against unwanted tilting.
7. Dismantle the clamp (8).
8. Carefully lift out the actuator and place it to the side.
☞
Note:
During disassembly of the lifting rod the connection
cable is on the left-hand side, in driving direction.
Mounting the lifting rod
Assembly of the lifting rod is done analogue in reverse or-
der to chapter
☞
Make sure that the collar (9) fits correctly and that dis-
tancers are located underneath the lower clamp (8).
!
Attention:
It is to be observed that the cables are not damaged or
bent when being placed (10).
☞
Otherwise danger of fire through short circuits!
☞
Previously removed cable ties are to be replaced.
☞
Then carry out a function test.
☞