METROTEC U15 Series Скачать руководство пользователя страница 7

METROTEC 

 

 

 

3

 

Introduction 

3.1 

Measuring Principle 

 

Oxygen  measurement  units  are  designed  to  process  signals  transmitted  from  an 
oxygen sensor constructed of stabilized zirconium oxide. Zirconium oxide, a ceramic 
material  that  is  also  spoken  of  as  a  solid-state  electrolyte,  acts  as  an  excellent 
oxygen-ion conductor when at a high temperature. 
Within  certain  temperature  limits,  that  depend  on  the  doping  of  the  material 
concerned,  such  ion  conductors  are  able  to fill  empty  spaces  in  their  crystal    lattice 
with  oxygen  ions.  The  oxygen  ions  occur  against  an  electrically  conductive  surface 
that is generally of platinum. 
The concentration of oxygen in a measurement gas is thus decisive for the extent of 
oxygen activity, and thus for the number of oxygen ions. 
An  oxygen  sensor  consists  essentially  of  a  solid-state  electrolyte  with  a  contact 
surface on both sides. 
One  side  of  the  electrolyte  is  in  contact  with  a  reference  gas  such  as  air,  and  the 
other  with  the  gas  whose  oxygen  content  is  to  be  measured.  The  mechanical 
construction  of  the  sensor  prevents  contact  between  the  two  gases  so  that  there  is 
no risk of their being intermixed. 
Depending  on  the  application  concerned,  heated  or  unheated  sensors  are  used. 
Unheated sensors are generally used in furnaces while heated sensors are used for 
applications where the gas to be measured is at  a temperature of less than around 
600  degrees  Celsius  (the  measurement  principle  necessitates  the  sensor  being 
maintained at a temperature of not less than 500 - 650 degrees Celsius). 
Heated  sensors  are  maintained  at  a  set  temperature  by  an  electronic  temperature 
regulator  that  forms  part  of  the  electronic  control  unit.  The  temperature  of  both 
heated and unheated sensors as measured by the electronic control is an important 
parameter  for  inclusion  in  the  calculation  of  the  oxygen  content  (oxygen  partial 
pressure) in accordance with the following equation:      

 

EMF

R T

F

P

P

4

1

2

ln(

)

 

 
whereby: 
 

R  =  8.31J/mol K 

T  =  Temperature in Kelvin 
F  =  96493 As/mol 

P1  =  Oxygen  partial  pressure  on  the  reference  side  with 

0.20946 bar 

P2  =  Oxygen  partial  pressure  on  the  measurement  gas 

side 

EMF  =  Electromotive forcein Volts 

Содержание U15 Series

Страница 1: ...Oxygen Measuring and Control Device U15 Series Measuring module Version 1 6 METROTEC Systeme zur Messung und Regelung von Sauerstoff U15 DIGITAL ...

Страница 2: ... Equipment Act Directive 2014 53 EU This device complies with following standards EN 61010 1 EN 61000 6 4 EN 61000 6 2 EN 63000 EN 300220 2 Description of measures taken to assure compliance Quality management system DIN EN ISO 9001 2015 No 12 100 27736 TMS This declaration becomes invalid if changes are made without our consent Kirchheim Teck 27 11 2019 Place Date Signature 2019 METROTEC GmbH Hei...

Страница 3: ...n 9 5 2 Measurement 9 5 3 Switching the Measuring Module off 9 6 Measured Value Output 10 6 1 LIN Measured Value Output 10 6 2 LOG Measured Value Output 10 7 Configuration 11 7 1 Enabling Configuration 11 7 2 Correcting the Measured Value 11 7 3 Adjusting the Limit Value 12 7 4 Practical approach 13 7 5 Definition of Analog Outputs 13 7 6 Analog Interfaces 15 7 7 Digital Interfaces 15 8 App for An...

Страница 4: ...ensure that the power supplies concur with the details listed on the nameplate All of the covers needed to ensure that the unit cannot be touched when operating must always be fitted You must consider the effects of the overall operation and take the necessary precautions if the unit will be linked up with other equipment and or devices before you switch on Parts and surfaces will occasionally bec...

Страница 5: ...the occurance of leakage corrosion condensation etc damage could be done to the equipment and incorrect measurement results be indicated and so it is essential that all parts of equipment be regularly serviced The oxygen sensor and its accessories are subjected to thorough quality control in accordance with DIN ISO 9001 in the course of their manufacture and testing They must only be installated a...

Страница 6: ... u15 Digital Designation Sensor Output Alarm Supply U15 Digital Extractive 1 2 230VAC U15 Digital 2 Extractive 2 2 230VAC U15 Digital 24 Extractive 1 2 24VDC U15 Digital 224 Extractive 2 2 24VDC U15 Digital C Inline 1 2 230VAC U15 Digital 2C Inline 2 2 230VAC U15 Digital C24 Inline 1 2 24VDC U15 Digital 2C24 Inline 2 2 24VDC ...

Страница 7: ...se oxygen content is to be measured The mechanical construction of the sensor prevents contact between the two gases so that there is no risk of their being intermixed Depending on the application concerned heated or unheated sensors are used Unheated sensors are generally used in furnaces while heated sensors are used for applications where the gas to be measured is at a temperature of less than ...

Страница 8: ...n order to implement extensive changes and to display measured values the METROTEC U15 App for Android Smartphones is required 3 3 Sensor Depending on the version the measuring module requires a separate extractive oxygen sensor or an inline sensor to carry out its functions Connecting the sensor See section Connection Diagram as well as separate operating instructions of the sensor 4 General Layo...

Страница 9: ...d measuring value have exceeded the limit value Figure 1 Figure 2 Figure 3 Ab Bereitschaft liefert das Messmodul am mA Ausgang den übertragenen Messwert 5 2 Measurement After the starting routine the device is ready for use and can determine the oxygen content in gases This requires supplying the sample gas according to the sensor type Alternatively the sample gas contains the sensor See the opera...

Страница 10: ...easured values implicates the assignment of the measured values to either the 0 to 20 mA or the 4 to 20 mA output in logarithmic mode This assignment is permanent and can only be changed on location by means of the METROTEC U15 App for Android Smartphone The output can take on values between 100 and 10 30 The values represent the logarithm of the oxygen partial pressure This presentation helps to ...

Страница 11: ...isplay of left LED array No settings possible Adjustment mode for reference point Adjustment mode for Measured value Setting mode for Limit value Figure 1 Figure 2 Figure 3 Figure 4 LED flashes while value is being changed via T3 or T4 7 2 Correcting the Measured Value The current measured value can be corrected if required In this case it is advisable to make the adjustment after a stable measure...

Страница 12: ...is flashing When finished the setting is saved by pressing T2 which also serves to leave the setup menu All LEDs in the left array are off See Table 1 Figure 1 7 3 Adjusting the Limit Value The limit value is switched when the current measured value exceeds the current limit value Usually the limit value is within the programmed measuring range An alarm status is signalled at the bottom of the rig...

Страница 13: ... a commercially available portable mA measuring device Now the adjustments and settings described above can be carried out After finishing these settings the previously disconnected connections are restored 7 5 Definition of Analog Outputs Switching the analog output from 4 to 20 mA to 0 to 20 mA occurs by means of a jumper between the terminals 12 and 13 Should there be a 2nd analog output it wil...

Страница 14: ...A Oxygen 0 5 10 15 20 0 5 10 15 20 mA Oxygen Switching the analog output to linear logarithmic occurs by means of a jumper between the terminals 16 and 17 Should there be a 2nd analog output it will also be switched over in the process Example for logarithmic output 0 to 20 mA at 0 01 100 oxygen ...

Страница 15: ...id Smartphone can be connected up As soon as the connection has been established the LED will light permanently Pressing T2 again will switch the Bluetooth connection and the green LED off An exact description is available in the METROTEC U15 App for Android Smartphone operating instructions 9 Specifications Measuring range 100 to 10 35 bar O2 Ambient temperature 0 to 45 degrees Celsius Measuring ...

Страница 16: ...mA Output 1 and 2 are electrically isolated from one another Connecting sensors Note Please refer to the operating instructions of the respective sensor Extraktive Sensoren Pin Function 1 Heating 2 Heating 3 Sensor 4 Sensor Inline Sensoren Pin Function 1 Thermocouple 2 Thermocouple 3 Sensor 4 Sensor ...

Страница 17: ...easuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 2 Terminal Function Definition 1 230 VAC Supply 2 230 VAC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 10 ...

Страница 18: ...measuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 2C Terminal Function Definition 1 230 VAC Supply 2 230 VAC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 1...

Страница 19: ...measuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 224 Terminal Function Definition 1 24 VDC Supply 2 24 VDC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 10...

Страница 20: ...measuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 2C24 Terminal Function Definition 1 24 VDC Supply 2 24 VDC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 1...

Отзывы: