USER MANUAL
SOLDERABILITY TESTER
METRONELEC
V2.0
–
10/2021
6 Avenue Eiffel - 78420 Carrières-sur-Seine
Tél : 01 30 15 20 00
Web :
http://www.metronelec.com| Mail : [email protected]
Page
36
sur
83
Zoom / scale (force, angle, time)
By a left click on the graph or from the menu Display you can switch from different
standards or scale, mix result or ask for an average.
You can personalize the scale in force or wetting angle, for this ask for the force or angle
scale and then User Scale, here is below the windows for the force, just fill the fields.
Validate / Invalidate / Open / Print / Delete / Average / Merge
Select the measurement with the mouse (left button), and click with the right button to
display the window here above.
Valid / Invalid
: Only valid curve will be displayed but invalid curves are still store on data.
Invalid curves are marked with a red cross, valid curve have a green mark.
Open
: open and display the graph if more than one curve are selected, the software open
one window by curve. (If you want just one graph, use the merge option).
: Print the current graph and or the table of values following the printing option.
Delete
: Erase from file data selected, be careful as there is no way to undelete.
Average
: Make an average curve of selected curves, calculate the average value and the
standard deviation in the table of values.