OMNIS Titrator
2.1001.0XX0
Product information
8.1001.8001EN / 2021-07-23
Страница 1: ...OMNIS Titrator 2 1001 0XX0 Product information 8 1001 8001EN 2021 07 23 ...
Страница 2: ......
Страница 3: ...Metrohm AG Ionenstrasse CH 9100 Herisau Switzerland 41 71 353 85 85 info metrohm com www metrohm com OMNIS Titrator 2 1001 0XX0 Product information 8 1001 8001EN 2021 07 23 ...
Страница 4: ...bove Disclaimer Deficiencies arising from circumstances that are not the responsibility of Metrohm such as improper storage or improper use etc are expressly excluded from the warranty Unauthorized modifications to the product e g conversions or attachments exclude any liability on the part of the manufacturer for resulting damage and its consequences Instructions and notes in the Metrohm product ...
Страница 5: ...substances 7 2 4 3 Danger from highly flammable substances 8 2 4 4 Danger from leaking liquids 8 2 4 5 Danger during transport of the product 9 2 5 Design of warning messages 9 2 6 Meaning of warning signs 10 3 Technical specifications 12 3 1 Ambient conditions 12 3 2 OMNIS Titrator Energy supply 12 3 3 Magnetic stirrer Energy supply 12 3 4 Measuring module Energy supply 13 3 5 OMNIS Titrator Dime...
Страница 6: ...7 3 14 Display specifications 17 3 15 Measuring Module Analog Display specifications 17 3 16 Measuring Module Digital Display specifications 17 3 17 Measuring Module Analog Measuring specifications 18 3 18 OMNIS Titrator Liquid Handling specifications 19 3 19 Magnetic stirrer Specifications 19 ...
Страница 7: ...able 1 Product versions Article number Designation Magnetic stirrer Function license 2 1001 0010 OMNIS Titrator without individual 2 1001 0020 OMNIS Titrator with individual 2 1001 0110 OMNIS Basic Titrator without Basic 2 1001 0120 OMNIS Basic Titrator with Basic 2 1001 0210 OMNIS Advanced Titrator without Advanced 2 1001 0220 OMNIS Advanced Titrator with Advanced 2 1001 0310 OMNIS Professional T...
Страница 8: ... number in accordance with GS1 standard 2 21 Serial number 3 240 Metrohm article number 4 Certification 5 Certification 6 Technical specifications 1 3 OMNIS Titrator Overview Figure 1 OMNIS Titrator Front 1 Lid 2 Space for measuring modules ...
Страница 9: ...le holder 2 Cable guide On the side to the right For Liquid Adapter cable 3 MSI connector MSI Metrohm Solution Identification Con nection socket for Liquid Adapter cable 4 Cable guide Bottom right 5 Interfaces 6 Cable guide Bottom left 7 Cable guide On the side to the left For rod stirrer cable or electrode cable ...
Страница 10: ...lways pro vides the current version of this document Further instructions leaflets release notes etc may be available depending on the product You can directly access the required information or the associated PDF document using the full text search function and filters 1 6 Accessories Up to date information on the scope of delivery and on optional accesso ries can be found on the Metrohm website ...
Страница 11: ...he Included parts tab click the link to download the PDF The PDF file with the accessories data is loaded NOTICE Metrohm recommends downloading the accessories list from the Internet and keeping it for reference purposes ...
Страница 12: ...ue to non observance of these limit values The EU declaration of conformity loses its validity as soon as modifications are carried out on the products and or the components 2 2 Responsibility of the operator The operator must ensure that basic regulations on occupational safety and accident prevention in chemical laboratories are observed The opera tor has the following responsibilities Instruct ...
Страница 13: ...r death To avoid danger from electrical potential observe the following Operate the product only if it is in perfect condition The housing must also be intact Only use the product with the covers fitted If covers are damaged or missing disconnect the product from the energy supply and contact the regional Metrohm service representative Protect live components e g power supply unit power cord conne...
Страница 14: ... 4 3 Danger from highly flammable substances Using highly flammable substances or gases may cause fires or explosions To avoid danger from highly flammable substances observe the following Avoid ignition sources Use protective grounding Use exhaust equipment 2 4 4 Danger from leaking liquids Leaking liquids may cause injuries and may damage the product To avoid danger from leaking liquids observe ...
Страница 15: ...here are 4 hazard levels for warning messages The following signal words are used for classifying the hazard levels in warning messages DANGER indicates a hazardous situation which if not avoided will result in serious injury or death WARNING indicates a hazardous situation which if not avoided could result in serious injury or death CAUTION indicates a hazardous situation which if not avoided cou...
Страница 16: ...uences when not observing the notice A minor to moderate injury is probable Measures to avoid the danger 2 6 Meaning of warning signs This documentation uses the following warning signs Table 2 Warning sign according to ISO 7010 Warning sign Meaning General warning sign Warning of electrical voltage Warning of hand injuries Warning of sharp object Warning of hot surface Warning of biological hazar...
Страница 17: ...g Warning of flammable materials Warning of corrosive substances Warning of optical radiation Warning of laser beams Depending on the intended use of the product the corresponding warn ing sign stickers must be placed on the product ...
Страница 18: ...elative humidity non condensing Storage 5 to 45 C 3 2 OMNIS Titrator Energy supply Nominal voltage range 100 VAC 240 VAC Nominal frequency range 50 Hz 60 Hz Power consumption max 100 W Fuse protection Internal fuse 4 ATH cannot be replaced by the user 3 3 Magnetic stirrer Energy supply Nominal voltage 24 VDC internal ...
Страница 19: ...ive coupling 3 5 OMNIS Titrator Dimensions Measurements Width 142 mm Height 358 mm Depth Without magnetic stirrer 284 mm With magnetic stirrer 400 mm Weight Type Without magnetic stirrer 4 4 kg With magnetic stirrer 5 1 kg 3 6 Magnetic stirrer Dimensions Measurements Width 142 mm Height 70 mm Depth 116 mm Weight 700 g ...
Страница 20: ...ck panel AW 5754 H12 H22 aluminum coated Base 1 4301 high grade steel Enclosure PBT poly butylene tereph thalate Front foils PET poly ethylene tereph thalate mat IP degree of protection IP 40 3 9 Magnetic stirrer Housing Materials Cover PBT poly butylene tereph thalate Base chromium steel Enclosure PBT poly butylene tereph thalate Front foils PET poly butylene tereph thalate mat IP degree of prote...
Страница 21: ...Connectors specifications Energy supply via power connection Socket IEC 60320 type C14 10 A Power cord Length max 2 m Number of conductors 3 with protective ground Conductor cross section min 0 75 mm2 18 AWG Plug Instrument side IEC 60320 type C13 10 A Building side country specific MDL Metrohm Device Link 4 connectors HID Human Interactive Device MSI Metrohm Solution Identifica tion LAN Local Are...
Страница 22: ... cyl inder unit 3 12 Measuring Module Analog Connectors specifica tions Measuring inputs INPUT 1 Socket round plug 7 pin size 0 45 Potentiometric pH ISE Redox measuring input for potentiometric electro des Temperature Temp measuring input for temperature sensors of the Pt1000 or NTC type for automatic tempera ture compensation Polarizer Pol measuring input for polarizable electrodes INPUT 2 Socket...
Страница 23: ...ox potentiometric differen tial measurement with respect to REF 3 13 Measuring Module Digital Connector specifications Connection socket for a digital elec trode Type round plug 6 pin size 0 60 3 14 Display specifications Status display LED multi colored 3 15 Measuring Module Analog Display specifications Status display LED green red 3 16 Measuring Module Digital Display specifications Status disp...
Страница 24: ... 10 12 A Temperature Pt1000 Measuring range 150 to 250 C Resolution approx 0 002 C Measuring accuracy 0 4 C in the measuring range 20 0 to 150 0 C Polarizer Ipol DC Polarization current 200 0 to 200 0 µA adjustable in 0 5 µA steps Measuring range 2 400 to 2 400 mV Measuring resolution 0 1 mV Potentiometric differential measure ment Measuring range 2 400 to 2 400 mV Measuring resolution 1 56 µV Mea...
Страница 25: ...sing resolution 100 000 steps per cylinder vol ume Dosing accuracy 0 01 typically of the dosed volume with respect to the cylinder volume 3 19 Magnetic stirrer Specifications OMNIS product version with attached magnetic stirrer Adjustment range for rotational speed 1 to 15 Rotation in counter clockwise direction seen from above 1 to 15 Rotation in clockwise direction seen from above Rotational spe...