![Metrohm 940 Professional IC Vario TWO Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/metrohm/940-professional-ic-vario-two/940-professional-ic-vario-two_manual_1779626065.webp)
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance
940 Professional IC Vario TWO (2.940.2100)
■■■■■■■■
57
3 Reinserting the inlet valve into the pump head
■
Insert the inlet valve into the inlet valve holder (the seal must be
visible) (
1
).
■
Screw the inlet valve holder up into the pump head (
2
) and
tighten it firmly by hand and then retighten it one additional
¼
turn using the adjustable wrench (
3
).
■
Refasten the connection capillary to the connection of the eluent
aspiration tubing
.
Removing the pump head
Accessories
For this step, you need the following accessories:
You can find these parts in the accessory kit:
Vario/Flex Basic
(6.5000.000).
■
4 mm hex key (6.2621.030)
Removing the pump head
Prerequisites:
■
Is the high-pressure pump switched off?
■
Has the pressure been released?
■
Is the instrument switched off?