3 Installation
892 Professional Rancimat
■■■■■■■■
17
Figure 8
Mounting the air tube: correct - incorrect
WARNING
The foam barrier can melt if it projects too deeply into the heating
block.
Ensure that the foam barrier is
at least 7 cm
above the base of the
reaction vessel.
Figure 9
Mounting the foam barrier
3.2.4
Inserting vessels / Establishing tubing connections
After you have assembled the reaction and measuring vessels, insert them
in the 892 Professional Rancimat and establish the tubing connections
1 Inserting the measuring vessel
■
Fill distilled water into the measuring vessel.
■
Place the measuring vessel cover onto the measuring vessel.
■
Insert the measuring vessel into the openings provided on the 892
Professional Rancimat. While doing so, carefully guide the con-
nector plug into the electrode connector.
■
Connect the white silicone tubing to the M8/olive tubing adapter
of the measuring vessel cover.