5 Operation and maintenance
884 Professional VA
■■■■■■■■
67
5.2
Performing a Dummy cell test
Should you note any problems or irregularities during or after determina-
tions (e.g. noisy curves), you can perform a dummy cell test. In this test,
the hardware and the electrode cables up to the measuring head are
checked; this allows potential problems to be localized or excluded. The
electrodes are not tested with the
Dummy cell test
.
Proceed as follows:
1
Load the
Dummy cell test
method template in
viva
.
2
Plug the three electrode cables onto the contacts of the dummy cell
(6.2813.050).
Observe the correct assignment of the electrode cables (WE, RE and
AE).
3
Place the measuring head cover and the stopper
onto the
measuring head.
4
Start the determination; at the prompt, select the name of the 884
Professional VA being used from the list and wait until the measure-
ment is completed.
If the curve meets the following conditions, then defects in the hard-
ware or the electrode cables can be ruled out:
■
The curve must be linear.
■
At –2.0 V, the current should be –1.67 mA ± 0.33 mA.
■
At +2.0 V, the current should be +1.67 mA ± 0.33 mA.