![Metrohm 863 Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/metrohm/863/863_manual_1779594057.webp)
■■■■■■■■■■■■■■■■■■■■■■
6 System settings
863 Compact Autosampler
■■■■■■■■
49
Files
Exported methods are stored in this directory.
The directory is created as soon as a backup is
exported for the first time.
Only methods located in this directory can be
imported.
6.3
Lift settings (Lift)
Menu
▶
System
▶
Lift
Work position
The working height of the lift can be set to the desired value. This is car-
ried out by directly operating the lift.
Three functions can be selected from the function bar with
[
⇦
]
and
[
⇨
]
and then executed by pressing
[OK]
:
■
Work pos.
moves the lift to the current working height.
■
Up
moves the lift 6 mm upward.
■
Down
moves the lift 6 mm downward.
When this dialog page is closed, the respective current lift position will be
applied as
Work position
.
Input range
0 - 126 mm (Increment: 6)
Default value
60 mm
The lift can only be moved in steps of 6 mm. If a more accurate lift posi-
tion setting is required, the needle position can be fine aligned.
Adjusting the needle position
First move the lift to working height.
1 Opening the safety shield
Loosen the lower two fastening screws of the safety shield using the
hex key provided, and tip up the safety shield.
2 Loosening the nut
Slightly loosen the nut below the lift head using a wrench.
Содержание 863
Страница 1: ...863 Compact Autosampler Manual 8 863 8003EN 2019 12 02 ...
Страница 2: ......