2 Installation
2.4
Connection of external instruments
2.4.1 General
information
Before an external instrument is connected to the 830 IC Interface, the
830 IC Interface must always be switched off using mains switch
Each system of the two versions of the 830 IC Interface has four RS232
interfaces
or
for connection of a maximum of 8 external instru-
ments, an
or
interface for analog signals from a maximum of 2 de-
tectors and 8 remote output lines
or
for controlling external
instruments by making contacts or impulses (see
about the technical data of these interfaces is given in
For installation and startup of external instruments proceed as follows:
1 Switch off all instruments
•
Switch off 830 IC Interface and all external instruments using
the mains switch.
2 Connect instruments
•
Connect the instruments to the 830 IC Interface according to
the connection diagrams shown in
to
by
using the cables named in the diagrams. Other instrument
combinations can be set up by using these examples as a
guide
.
3 Switch on all instruments
•
Switch on 830 IC Interface and all external instruments using
the mains switch.
4 Instrument settings
Operation with the 830 IC Interface requires that the following
settings must be made:
•
Advanced IC instruments:
818 IC Pump
819 IC Detector
820 IC Separation Center
833 IC Liquid Handling Unit
These instruments are already pre-configured, no additional
settings must be made.
Also all other Metrohm IC instruments can be operated with the
830 IC Interface, for the settings of the instruments please refer
to the respective instructions for use.
•
732 IC Detector:
>CONFIG/RS settings
handshake: SWchar
830 IC Interface / 8.830.1003 Instructions for Use
10