4 Start-up
947 Professional UV/VIS Detector Vario (2.947.00x0)
■■■■■■■■
31
4.1.1
Operation with UV lamp and VIS lamp
NOTICE
Only follow the following steps if the instrument is in operation with
the UV lamp and the VIS lamp.
The lamp settings must be adjusted if the lamp was replaced or if
checking the intensity spectrum after first start-up displays a cut-off
Adjusting the UV lamp and the lamp settings
Prerequisites
■
The UV lamp and the VIS lamp are installed.
■
The instrument is switched on.
■
The instrument was configured in MagIC Net
(see 8.102.8004EN Man-
ual MagIC Net 3.3)
.
■
The UV lamp and the VIS lamp have been burning for 30 minutes.
■
The flow-through cell is clean.
■
The flow-through cell is rinsed with ultrapure water.
■
The flow-through cell is free of air bubbles.
1
Open the
Configuration
program part in MagIC Net.
Double-click on
947 UV/VIS Detector
in the device table.
The
Properties...
window opens.
2
On the
Detector
tab, click on
[Properties...]
to open the detector
settings.
3
Set the integration duration manually to 20.0 ms. Set the VIS inten-
sity level manually to 7.0. Confirm with
[Apply]
.