6.1 Removing the scTRACE Gold
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
6 Maintenance
6.1
Removing the scTRACE Gold
To prevent cross-contaminations between different applications or if the
electrode performance is bad, the scTRACE Gold electrode can be
replaced.
NOTICE
The scTRACE Gold electrode has a limited lifetime. If the performance
of the electrode cannot be recovered by the cleaning procedure descri-
bed in the respective application description or in
ing of the scTRACE Gold electrode (page 38)
, the electrode needs to be
replaced. Indications for an outdated electrode are:
■
Significantly lower sensitivity.
■
Higher or very noisy background current.
■
Unusual peak shape.
Figure 13
Remove scTRACE Gold electrode
1
Remove the measuring head, with transparent ring and scTRACE
Gold assembled from the measuring stand.