5 Troubleshooting
2.945.0010 945 Professional Detector Vario – Conductometry
■■■■■■■■
21
5 Troubleshooting
5.1
Problems and their solutions
Problem
Cause
Remedy
The pressure in the
system markedly
increases.
The conductivity detector is
blocked.
■
Shorten the capillary ends by a few millime-
ters.
■
Rinse the detector opposite the normal
flow direction.
Conductivity detec-
tor is not recognized
in the software
No connection.
■
Check the connection of the detector
cable.
■
Switch the instrument off and (after 15 sec-
onds) on again.