3 Installation
942 Extension Module Vario SeS/PP (2.942.0500)
■■■■■■■■
33
4 Putting the adsorption cartridge holder into the instrument
■
Place the adsorption cartridge holder with the CO
2
adsorption car-
tridge CW into the instrument's detector chamber.
3.7
Connecting an Extension Module
CAUTION
The 940 Professional IC Vario
has to be switched off
when connect-
ing the Extension Module!
Accessories
For this step you need the following accessories:
■
6.2156.060 cable Extension Module - Professional IC, 40 cm
or
■
6.2456.070 cable Extension Module - Professional IC, 1 m (optional
accessory)
The connection sockets are located on the rear of the instrument.
Connecting the Extension Module to the IC instrument
1
Plug the connecting cable (6.2156.060) into the
In
connection
socket on the Extension Module and tighten it in place.
2
Plug the other end of the connecting cable into the
Extension
Module
connection socket on the IC instrument and tighten it in
place.