3 Installation
942 Extension Module Vario SeS/PP (2.942.0500)
■■■■■■■■
11
CAUTION
Do not pinch capillaries or leak sensor cables
Capillaries are fed through the guide ducts between the base tray and
the instrument. Pinches in the leak sensor cable or the capillaries may
lead to malfunctions.
■
Unplug the leak sensor cable before you remove the base tray.
■
Remove all the capillaries from the capillary ducts before you remove
the base tray.
Removing the base tray
Prerequisites
■
The instrument is switched off.
■
The bottle holder is cleared.
■
All of the cable connections on the rear have been disconnected.
■
The capillaries are removed from the guide ducts between the instru-
ment and the base tray.
■
There are no loose parts in the instrument.
Accessories
■
3 mm hex key (6.2621.100)
1
2
3
Figure 4
Removing the base tray
1
Tilt the instrument sideways and lay it down flat.
2
Loosen the 4 cylinder screws with the 3 mm hex key. Remove the cyl-
inder screws and their washers.