8 Technical specifications
889 IC Sample Center
■■■■■■■■
49
8.3
Analytical characteristics
Injection modes
Full-Loop, Partial Loopfill, Pickup mode, PASA
™
(pressure-assisted sam-
ple aspiration)
Reproducibility
Relative standard deviation
≤
0.3% for the Full-Loop injection.
Relative standard deviation
≤
0.5% for the Partial Loopfill injection, injec-
tion volume > 10 µL.
Relative standard deviation
≤
1.0% for the Pickup injection, injection vol-
ume > 10 µL.
Memory effect
< 0.05 % with programmable needle rinsing
8.4
Programming
Injection modes
Full-Loop, Partial Loopfill, Pickup Mode
Injection volume
1…5000 µL (with 1 µL increment), depending on the system settings.
Max. injection vol-
ume
■
Full Loop = Sample loop volume
■
Partial Loopfill = 0.5 x Sample loop volume
■
Pickup = (Sample loop volume - 3 x Needle volume)/2
8.5
Interfaces
Output signal
1 programmable relay outlet as injection mark
Data interface
USB connector, type B (for the connection to a PC)
8.6
Options (pre-installed)
Cooling for the
sample rack
Built-in Peltier element
Range: 4 °C to 3 °C below ambient temperature
Air temperature in the sample vessel: 4 °C
±
2 °C (at the temperature
sensor)
(Temperature at 80% relative humidity and ambient temperature of
25°C)