6 Troubleshooting
889 IC Sample Center
■■■■■■■■
39
Problem
Cause
Remedy
If the error occurs again, please call the
Metrohm Service.
Error 297
The Home sensor of the rack
holder has unexpectedly been
activated.
Call the Metrohm Service.
Error 298
The rack is located in an unde-
fined position.
Carry out the function
[Reset device]
in
the Manual Control of the sample
changer.
If the error occurs again, please call the
Metrohm Service.
6.1.2 Needle unit
Problem
Cause
Remedy
Error 303
The needle is located in an
undefined horizontal position.
Carry out the function
[Reset device]
in
the Manual Control of the sample
changer.
If the error occurs again, please call the
Metrohm Service.
Error 304
The Home sensor for the hori-
zontal movement of the nee-
dle has not been activated.
1. Check whether the horizontal move-
ment of the needle is obstructed.
2. If possible, remove possible obstacles.
If the error occurs again, please call the
Metrohm Service.
Error 306
The Home sensor for the hori-
zontal movement of the nee-
dle has not been deactivated.
1. Check whether the horizontal move-
ment of the needle is obstructed.
2. If possible, remove possible obstacles.
If the error occurs again, please call the
Metrohm Service.
Error 307
The Home sensor for the hori-
zontal movement of the nee-
dle has unexpectedly been
activated.
Call the Metrohm Service.
Error 308
In order to reach the horizontal
Home position of the needle,
the expected number of steps
has not been carried out.
1. Check whether the horizontal move-
ment of the needle is obstructed.
2. If possible, remove possible obstacles.