3.3 Mounting the extension module below the IC instrument
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
872 Extension Module IC Pump
9 Mounting the drainage tubing
Reconnect the drainage tubing to the drainage tubing connector of
the bottle holder.
Possibly, a longer section of silicone tubing 6.186.020 must be cut to
fit and mounted
(see also the manual for the IC instrument)
.
3.3
Mounting the extension module below the IC instru-
ment
1 Switching off the IC instrument
Switch the IC instrument off and disconnect the mains cable.
2 Clearing the bottle holder
If there are bottles and other things on the bottle holder, remove
them.
3 Disconnecting all connections on the rear of the instrument
■
Disconnect the mains cable,
■
Disconnect the MSB cable,
■
Disconnect the USB cable,
■
Disconnect the leak sensor,
■
Remove the drainage tubings.
4 Removing the detector(s)
Disconnect the detector cable(s) and remove the detector(s) from the
IC instrument
(see manual for the IC instrument)
.
5 Removing the base tray
■
Tilt the IC instrument sideways and lay it down flat.
■
Loosen the cylinder screws with a 6.2621.100 3 mm hexagon
key.
■
Remove base tray.