2.3 Connecting the detector
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
2.850.9010 IC Conductivity Detector
2.3
Connecting the detector
Connecting the detector cable and the detector outlet
capillary
CAUTION
The instrument
must
be switched off when connecting a detector.
1 Connecting the detector
■
Connect the detector connection cable
to the detector con-
nection socket
Detector 1
.
NOTICE
Recommendation
: In AnCat systems with 2 detectors: anions to
Detector 1
, cations to
Detector 2
.
2 Connecting the detector outlet capillary
■
Guide the detector outlet capillary
into a sufficiently large
waste container and fasten it there.
NOTICE
The detector outlet capillary must be free of blockages (the mea-
suring cell is tested to 5 MPa = 50 bar backpressure).
NOTICE
The detector outlet capillary must
not
be shortened!
Shortening the detector outlet capillary leads to increased noise!