3.15 Sample degasser
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
3
4
5
2
3
6
1
2
Figure 25
Pulsation damper – Connection
1
Connection capillary
Connection to the inline filter.
2
Fastening screws
3
PEEK pressure screws, short
6.2744.070
4
Holder for pulsation damper
5
Pulsation damper 6.2620.150
6
Connection capillary
Connection to the injection valve.
3.15
Sample degasser
The sample degasser removes gas bubbles and disolved gases from the
sample. For degassing, the sample flows into a vacuum chamber through
a special fluoropolymer capillary.
Gas bubbles in the sample lead to poor reproducibility, as the quantity of
sample in the sample loop would not always be the same. Samples (con-
taining gas) should therefore be degassed before injection. For this the
sample is sucked through a degasser chamber before injection, whereby
any gas bubbles are automatically removed.
NOTE
When using the sample degasser, the rinsing time extends by at least 2
minutes.