Metrohm 2.803.0010 Скачать руководство пользователя страница 1

803 Ti Stand

Manual

8.803.8001EN / 2019-09-26

Содержание 2.803.0010

Страница 1: ...803 Ti Stand Manual 8 803 8001EN 2019 09 26 ...

Страница 2: ......

Страница 3: ...Metrohm AG CH 9100 Herisau Switzerland Phone 41 71 353 85 85 Fax 41 71 353 89 01 info metrohm com www metrohm com 803 Ti Stand Manual 8 803 8001EN 2019 09 26 ...

Страница 4: ...techcom metrohm com This documentation is protected by copyright All rights reserved This documentation has been prepared with great care However errors can never be entirely ruled out Please send comments regarding possible errors to the address above ...

Страница 5: ...lvents and chemicals 3 1 3 5 Recycling and disposal 4 2 Overview of the instrument 5 3 Installation 7 3 1 Setting up the instrument 7 3 1 1 Packaging 7 3 1 2 Checks 7 3 1 3 Location 7 3 2 Mounting the 803 Ti Stand 7 3 3 Connecting the 803 Ti Stand 10 3 4 Mounting the accessories 12 4 Troubleshooting 15 4 1 Problems 15 5 Technical specifications 16 5 1 Rotational speed 16 5 2 Pump 16 5 3 Power supp...

Страница 6: ...gure 1 Front 803 Ti Stand 5 Figure 2 Rear 803 Ti Stand 6 Figure 3 Connecting the 803 Ti Stand 10 Figure 4 Connecting the Dosino to the 803 Ti Stand 11 Figure 5 803 Ti Stand with volumetric titration vessel and equipment for auto matic reagent exchange 12 ...

Страница 7: ... the documentation CAUTION Please read through this documentation carefully before putting the instrument into operation The documentation contains information and warnings which the user must follow in order to ensure safe opera tion of the instrument 1 2 1 Symbols and conventions The following symbols and styles are used in this documentation Cross reference to figure legend The first number ref...

Страница 8: ...eneral notes on safety WARNING Operate this instrument only according to the information contained in this documentation This instrument left the factory in a flawless state in terms of technical safety To maintain this state and ensure non hazardous operation of the instrument the following instructions must be observed carefully 1 3 2 Electrical safety The electrical safety when working with the...

Страница 9: ...er socket before you set up or disconnect electrical plug connections at the rear of the instrument 1 3 3 Working with liquids CAUTION Periodically check all system connections for leaks Observe the relevant regulations in respect to working with flammable and or toxic fluids and their disposal 1 3 4 Flammable solvents and chemicals WARNING All relevant safety measures are to be observed when work...

Страница 10: ...EU WEEE Waste Electrical and Electronic Equipment The correct disposal of your old instrument will help to prevent negative effects on the environment and public health More details about the disposal of your old instrument can be obtained from your local authorities from waste disposal companies or from your local dealer ...

Страница 11: ...ump and magnetic stirrer 3 LED function display Lights up when the stirrer is switched on 4 Key Pressing the key aspirates air out of the aspi ration bottle The vacuum in the suction bot tle suctions the liquid out of the KF titration cell and into the suction bottle 5 Key Pressing the key pumps air into the solvent bottle The overpressure in the solvent bottle pushes solvent into the KF titration...

Страница 12: ...Stand 1 Connection nipple for PVC tubing For aspirating the titration vessel content 2 Connection nipple for PVC tubing For aspirating solvent 3 MSB connector For connecting additional devices e g Dos ino 4 MSB connection cable For connecting to an analysis device ...

Страница 13: ... note 3 1 3 Location The instrument has been developed for operation indoors and may not be used in explosive environments Place the instrument in a location of the laboratory which is suitable for operation and free of vibrations and which provides protection against corrosive atmosphere and contamination by chemicals The instrument should be protected against excessive temperature fluctua tions ...

Страница 14: ...the opening on the bottom of the KF titration stand onto the hexagon screw in the stand plate 3 Slide the 6 2016 070 support rod from above into the opening of the KF titration stand intended for this purpose up to the hexagon screw and tighten 6 2016 070 ...

Страница 15: ...t the 6 1414 030 vessel lid of the KF titration cell to the sup port rod Keep the locking lever pressed down until it can be released at the desired position 3 Fasten the 6 1415 220 or 6 1415 250 titration vessel with a 6 1903 020 or 6 1903 030 stirring bar inside on the vessel lid Fold back the holding bracket upwards while doing so The markings on the vessel lid and on the plastic ring must be a...

Страница 16: ...now be fixed by readjusting the clamping ring Once the height of the KF titration cell has been adjusted correctly the entire cell can be raised and swiveled to the right as required by pressing the locking lever 3 3 Connecting the 803 Ti Stand Connect the titration stand as follows 1 Exit the control software 2 Connect the connection cable of the titration stand to one of the sockets marked with ...

Страница 17: ... 3 Installation 803 Ti Stand 11 MSB Waste Solvent Figure 4 Connecting the Dosino to the 803 Ti Stand 4 Start the control software ...

Страница 18: ...ng out of reagent bottles with GL 45 thread 7 Cannula 6 1819 050 8 Amber glass bottle 6 1608 030 Waste bottle 9 PVC tubing 6 1801 120 10 Aspiration tip 6 1543 120 11 Connection nipple for PVC tubing For aspirating the titration vessel content 12 PTFE tubing 6 1805 200 13 Transfer tip 6 1543 110 14 Connection nipple for PVC tubing For aspirating solvent 15 PVC tubing 6 1801 120 16 Threaded stopper ...

Страница 19: ...tle attachment onto the 6 1608 030 waste bottle 4 Cut the 6 1801 120 PVC tubing into two halves and use one to aspi rate 5 Fasten the one end of the aspiration tubing to the tubing adapter the other to the connection nipple for the waste on the rear of the titration stand 6 Insert the 6 1819 050 cannula through the 6 1446 090 stopper into the waste bottle 7 Screw the 6 1805 200 PTFE tubing into th...

Страница 20: ...r tube 4 Secure the adsorber tube with the 6 2023 020 SGJ clip 5 Insert the 6 1819 030 cannula into the last open hole with medium sized diameter of the 6 1602 105 bottle attachment 6 Screw the 6 1602 105 bottle attachment onto the solvent bottle Instead of the 6 1608 023 amber glass bottle other reagent bottles with GL45 thread can be used e g RIEDEL DE HAEN 1 liter BAKER For other bottles anothe...

Страница 21: ...s no liquid The add aspirate key is not pressed tightly enough It is usually the case that the bottle attach ments are not screwed tightly enough to the reagent and waste bottles Check all connections for leak tightness Check whether the 803 Ti Stand has been set up correctly The solvent flows on into the titration vessel after the addition without the add key being pressed The solvent bottle is n...

Страница 22: ... Increase of rota tional speed per step 115 125 U min 5 2 Pump Add 600 mL min at 25 C standard accessories Aspirate 400 mL min at 25 C standard accessories 5 3 Power supply Voltage 12 V 12 V 5 V Power consump tion 4 W Fuse Electronic overload protection 5 4 Ambient temperature Nominal function range 5 45 C at a maximum of 85 humidity Storage 20 60 C Transport 40 60 C ...

Страница 23: ... Technical specifications 803 Ti Stand 17 5 5 Dimensions Width 106 mm Height without stand 101 mm Height with stand 412 mm Depth 220 mm Weight 1100 g 5 6 Material of housing Polybutylene terephthalate PBT ...

Страница 24: ...r Internet browser 2 Enter the article number e g 803 into the search field The search result is displayed 3 Click on the product Detailed information regarding the product is shown on various tabs 4 On the Included parts tab click on Download the PDF The PDF file with the accessories data is created NOTICE Once you have received your new product we recommend download ing the accessories list from...