4 Operation
873 Biodiesel Rancimat
■■■■■■■■
113
3 Set second tangent
■
Keep the
[Shift]
key and
left mouse button pressed down
to
set the second tangent.
4 Adjust the position of the tangents
■
Position the cursor on a black square on the tangent.
The cursor assumes the form of a cross hair.
■
Keep the
[Ctrl]
key and the
left mouse button pressed down
to move the point to the required position.
After you have adjusted the position of both tangents, you can read
the value of the new induction time both in the dialog window
Live
graph
at the top right beside the red line and also in the dialog window
Reevaluation
under
Manual
.
5 Save changes
■
In the dialog window
Live graph
click on the button
[Save]
.
The new induction time is accepted in the determination overview.
Deleting manually set tangents
Proceed as follows to delete the previously set tangents:
1 Delete tangents
■
Click on the symbol
.
Resetting all reevaluations
1 Reset induction and stability time to the original value
■
Click on the
[Reset]
button.
4.7.4
Extrapolation
Extrapolation can be used to extrapolate the results measured at various
temperatures to the required target temperature. In addition, the
factor
for
converting the
induction time
to the
standard time
can be determined.