INSTRUCTIONS

MX 67

DIGITAL TESTER

(WITH BACK LIGHT)

SAFETY  INFORMATION

The following safety information must be
observed to insure maximum personal safety
during the operation of this meter:
1. Do not use if the meter or test leads look

damaged, or if you suspect that the meter
is not operating properly.

2. This meter is not recommended for high

voltage industrial use; for example, not
for measurements of 440 VAC or 600
VAC industrial power mains. The unit is
intended for use with low energy circuits
to 600V AC/DC or high energy circuit to
250 VAC or DC.

3. Use caution when working above 60V dc

or 30V ac rms. Such voltages pose a
shock hazard.

4. When using the probes, keep your fingers

behind the finger guards on the probes.

5. Measuring voltage which exceeds the

limits of the meter may cause damage
and expose the operator to a shock haz-
ard. Always respect the voltage limits as
stated on the meter.

6. If the equipment is used in a manner not

specified by the manufacturer, the pro-
tection provided by the equipment may
be impaired.

7. Fused test leads in accordance with

H.S.E. guidance note GS38 should be
used in high energy circuits.

SPECIFICATIONS

Display: 3½ digit liquid crystal display (LCD) with a maximum reading of

3200.

Analog bar graph: 32 segments with measurements 12 times per second.
Polarity: Automatic, (-) negative polarity indication.
Overrange: "OL" mark indication.
Low battery indication: The "

" is displayed when the battery voltage

drops below the operating level.

Auto power-down: After 10 mins, Back-Light after 30 secs approx.
Measurement rate: 2 times per second, nominal.
Operating environment: 0°C to 40°C at <70% relative humidity.
Storage environment: -20°C to 60°C at <80% relative humidity.
Safety: According to EN61010-1 protection class II overvoltage category

(CAT II 600V) pollution degree 2.

Power: 4 pcs 1.5V (AAA size) UM-4 R03.
Battery life: 500 hours typical with alkaline.
Dimensions: 170mm (H) x 44mm (W) x 40mm (D).
Weight: 140g including batteries.

VOLTAGE

(40 / 500Hz)

Range

DC Accuracy

AC Accuracy

Input impedance

Maximum input

320mV

0.25%rdg+1dgt

No range

>1000ΜΩ

600VDC or ACrms

3.2V

1.0%rdg+4dgts

11ΜΩ

32V

10ΜΩ

320V

600V

OHM

Range

Resolution

Accuracy

Test Current

Input protection

320Ω

0.1Ω

0.3%rdg+3dgts

< 0.7mA

500VDC or ACrms

3.2ΚΩ

1Ω

0.3%rdg+1dgt

< 0.13mA

32ΚΩ

10Ω

< 13µA

320ΚΩ

100Ω

< 1.3µA

3.2ΜΩ

1ΚΩ

1.2%rdg+1dgt

< 0.13µA

32ΜΩ

10ΚΩ

3.0%rdg+1dgt

DIODE Test

0-2.0

1mV

3.0%rdg+3dgts

0.8mA±0.6mA

500VDC or ACrms

CONTINUITY

Check

Range

Resolution

Audible indication

Test Current

Input protection

320Ω

0.1Ω

<20Ω±5Ω

<0.7mA

500VDC or ACrms

OPERATION

Before taking any measurements, read the Safety Information Section.
Always examine the instrument for damage, contamination (excessive
dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed
insulation. If any abnormal conditions exist do not attempt to make any
measurements.

Autoranging

The meter defaults to autorange when you turn it on. In autorange, the
meter selects the range automatically.

MX 67-MOD (NSN 6625-99-0834221)