PRESS TO RE-POWER
TESTER
MX67
METRIX
ELECTRONICS
OPERATING
INSTRUCTIONS
MX 67
DIGITAL TESTER
(WITH BACK LIGHT)
SAFETY INFORMATION
The following safety information must be
observed to insure maximum personal safety
during the operation of this meter:
1. Do not use if the meter or test leads look
damaged, or if you suspect that the meter
is not operating properly.
2. This meter is not recommended for high
voltage industrial use; for example, not
for measurements of 440 VAC or 600
VAC industrial power mains. The unit is
intended for use with low energy circuits
to 600V AC/DC or high energy circuit to
250 VAC or DC.
3. Use caution when working above 60V dc
or 30V ac rms. Such voltages pose a
shock hazard.
4. When using the probes, keep your fingers
behind the finger guards on the probes.
5. Measuring voltage which exceeds the
limits of the meter may cause damage
and expose the operator to a shock haz-
ard. Always respect the voltage limits as
stated on the meter.
6. If the equipment is used in a manner not
specified by the manufacturer, the pro-
tection provided by the equipment may
be impaired.
7. Fused test leads in accordance with
H.S.E. guidance note GS38 should be
used in high energy circuits.
SPECIFICATIONS
Display: 3½ digit liquid crystal display (LCD) with a maximum reading of
3200.
Analog bar graph: 32 segments with measurements 12 times per second.
Polarity: Automatic, (-) negative polarity indication.
Overrange: "OL" mark indication.
Low battery indication: The "
" is displayed when the battery voltage
drops below the operating level.
Auto power-down: After 10 mins, Back-Light after 30 secs approx.
Measurement rate: 2 times per second, nominal.
Operating environment: 0°C to 40°C at <70% relative humidity.
Storage environment: -20°C to 60°C at <80% relative humidity.
Safety: According to EN61010-1 protection class II overvoltage category
(CAT II 600V) pollution degree 2.
Power: 4 pcs 1.5V (AAA size) UM-4 R03.
Battery life: 500 hours typical with alkaline.
Dimensions: 170mm (H) x 44mm (W) x 40mm (D).
Weight: 140g including batteries.
VOLTAGE
(40 / 500Hz)
Range
DC Accuracy
AC Accuracy
Input impedance
Maximum input
320mV
0.25%rdg+1dgt
No range
>1000ΜΩ
600VDC or ACrms
3.2V
1.0%rdg+4dgts
11ΜΩ
32V
10ΜΩ
320V
600V
OHM
Range
Resolution
Accuracy
Test Current
Input protection
320Ω
0.1Ω
0.3%rdg+3dgts
< 0.7mA
500VDC or ACrms
3.2ΚΩ
1Ω
0.3%rdg+1dgt
< 0.13mA
32ΚΩ
10Ω
< 13µA
320ΚΩ
100Ω
< 1.3µA
3.2ΜΩ
1ΚΩ
1.2%rdg+1dgt
< 0.13µA
32ΜΩ
10ΚΩ
3.0%rdg+1dgt
DIODE Test
0-2.0
1mV
3.0%rdg+3dgts
0.8mA±0.6mA
500VDC or ACrms
CONTINUITY
Check
Range
Resolution
Audible indication
Test Current
Input protection
320Ω
0.1Ω
<20Ω±5Ω
<0.7mA
500VDC or ACrms
OPERATION
Before taking any measurements, read the Safety Information Section.
Always examine the instrument for damage, contamination (excessive
dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed
insulation. If any abnormal conditions exist do not attempt to make any
measurements.
Autoranging
The meter defaults to autorange when you turn it on. In autorange, the
meter selects the range automatically.
MX 67-MOD (NSN 6625-99-0834221)