Appendix E
147
E.4
Creating / modifying an Auto Sequence®
If creating a new Auto Sequence
®
from scratch, the first step (Header) and the last step (Result)
are offered by default. Measurement steps are inserted by the user.
Options:
Adding a measurement step
By double clicking on a Single test a new
measurement step will appear as the last of
measurement steps. It can also be dragged and
dropped on the appropriate position in the Auto
Sequence
®
.
Adding flow commands
Selected flow command can be dragged from the list
of Flow commands and dropped on the appropriate
place in any Auto Sequence
®
step.
Changing position of flow command
within measurement step
By a click on an element and use of
,
keys.
Viewing / changing parameters of
flow commands or single tests.
By a double click on the element.
Setting number of measurement
step repetitions
By setting a number in the
field.
Note:
Up to 999 step repetitions can be set.
Right click on the selected measurement step / flow command:
Copy
– Paste before
A measurement step / flow command can be copied and
pasted above selected location on the same or on another
Auto Sequence
®
.
Copy
– Paste after
A measurement step / flow command can be copied and
pasted under selected location on the same or on another
Auto Sequence
®
.
Delete
Deletes the selected measurement step / flow command.
Содержание MI 3394 CE MultiTesterXA
Страница 1: ...CE MultiTesterXA MI 3394 Instruction manual Ver 3 4 13 Code no 20 752 432...
Страница 80: ...MI 3394 CE MultiTesterXA Single tests 79 Figure 6 19 Examples of Continuity measurement results...
Страница 103: ...MI 3394 CE MultiTesterXA Single tests 102 Figure 6 57 Examples of Power measurement results...
Страница 110: ...MI 3394 CE MultiTesterXA Single tests 109 Save results optional Figure 6 69 Examples of Functional Inspection results...