MI 3202 GigaOhm 5 kV
Measurements
17
Set-up test voltage
for Insulation
Resistance (
Fig.13
):
Adjust test voltage using the
and
keys
.
Fig. 13. Set-up test voltage in Insulation
Resistance measurement
Legend of displayed symbols:
INSULATION RESISTANCE
Name of selected function
Unominal
250V
Set test voltage
Set-up Insulation resistance limit
value
for Insulation Resistance (
Fig.13
):
-
Press the key
,
-
Adjust the limit value using the
keys
and
,
-
Press the key
again or the key
START to return to into insulation
resistance measurement menu.
Fig. 13a. Set-up Insulation Resistance limit
value
Legend of displayed symbols:
INSULATION RESISTANCE
Name of selected function
Resistance value
1 M
Selected insulation resistance limit
Insulation resistance limit selection range: [10 k
.. 200 M
]
No
limit
selected
indication: –
–
–
Warning!
Refer to Warnings chapter for safety precautions!
5.4 Voltage measurement
Select this function by pressing the
V
key. Voltage measurement is active immediately
after entering the function. See
Fig. 14
below.