MI 3102(H) BT
Eurotest
XE
Technical specifications
100
10.11
Power tests
Measurement characteristics
Function symbols
Class according to
IEC 61557-12
Measuring range
P
2.5
5 % ÷ 100 % I
Nom
(1)
Q
2.5
5 % ÷ 100 % I
Nom
(1)
S
2.5
5 % ÷ 100 % I
Nom
(1)
PF
1
-
1 ÷ 1
f
0.05
40 Hz ÷ 60 Hz
I, INom
1.5
5 % ÷ 100 % I
Nom
U
1.5
110 V ÷ 500 V
Uh
n
2.5
0 % ÷ 20 % U
Nom
THDu
2.5
0 % ÷ 20 % U
Nom
Ih
n
2.5
0 % ÷ 100 % I
Nom
THDi
2.5
0 % ÷ 100 % I
Nom
(1)
– I
Nom
depends on set current sensor type and selected current range:
A 1018, A1019 (20 A),
A 1391 (40 A or 300 A)
Note:
Error of external voltage and current transducers is not considered in this specification.
Function
Measuring range
Power (P, S, Q)
0.00 W (VA, Var) ÷ 99.9 kW (kVA, kVar)
Power factor
-
1.00 ÷ 1.00
Voltage harmonics
0.1 V ÷ 500 V
Voltage THD
0.1 % ÷ 99.9 %
Current harmonics and Current THD
0.00 A ÷ 199.9 A
Note: