MI 3155
Eurotest
XD
Appendix A
– Profile notes
218
A.2
Profile Hungary (ATAG)
Fuse type gR added to the fuse tables. Refer to
Fuse tables guide
for detailed information on
fuse data.
New Single test function
Visual Test
added.
Figure A.1: Visual Test menu
Measurement parameters / limits
Protection type
Protection type
[No, Automatic disconnection, Class II,
Electrical separation, SELV,PELV]
Measurement procedure
Enter the
Visual Test
function.
Set test parameters / limits.
Perform the visual inspection on tested object.
Use
to select PASS / FAIL / NO STATUS indication.
Save results (optional).
Figure A.2: Examples of Visual Test result
Added tests with multiplication factor 2.
Modification of RCD Auto test procedure