CS 2890 Power Simulator / Calibrator
General setup
20
To ensure correct measurement, it is essential that the CS 2890 instrument is regularly
calibrated. If used continuously on a daily basis, a
three-month calibration period
is
recommended, otherwise
annual calibration
is sufficient.
3.2.2 Adjustment of voltage output
1. Adjustment of CS 2890 voltage (between terminals L1
– L2) by
using
Calibration certificate
:
Figure 11: CS 2890 Calibration certificate
– voltage output
Selection of 5V on the CS 2890 generates 5,00450V on the CS 2890 output (value from
the calibration certificate), so it is necessary to adjust the voltage correction of 5,0045V.
Press ADJUST (1.)
Press
to Enter the Adjustment menu
Press Enter and
enter the corrected value of 5,0045V and confirm the value by pressing Enter key
Figure 12: Adjustment procedure using values from the calibration certificate