MI 2142 AlphaPAT Autotest modes
33
7.1.5 How to view autotest shortcut sequence measurement parameters
After pressing ENTER key in
Autotest shortcut menu
the measurement parameters of
the selected autotest sequence are detailed.
Autotest shortcut view menu
By using and cursor keys the parameters of the selected measurement can be
viewed.
The following measurement parameters can be viewed in all measurements:
measurement mode,
output test voltage or current (except in visual test),
pass level (except visual test), and
measurement duration (except in visual test).
In
disable
measurement mode the selected measurement is skipped.
In
single
measurement mode one measurement will be performed.
In
continuous
measurement mode up to five repetitive measurements can be
performed.
Press ESC key to return to the
Autotest shortcut menu
.
7.1.6 How to start autotest shortcut sequence
Pressing START key in Autotest shortcut menu starts the pre-defined autotest
sequence. Refer to chapter
7.4 Performing autotest sequences
for more information.
7.2 Autotest custom menu
Autotest custom menu allows setting and editing user-defined autotest procedures. 20
autotest sequences with all measurements disabled are added to the list by default and
can be set to user needs. They can be restored to default settings by selecting
ORIGINAL SETTINGS in
Setup menu
. Up to 50 custom autotest sequences can be
pre-programmed in this autotest mode.
Use and cursor keys in
Main menu
to select
Autotest custom
and press ENTER
key to confirm. The
Autotest custom menu
is displayed.
Autotest custom menu
Note:
Up to 50 custom autotest sequences can be pre-programmed! If you want to
save more than 50 autotests, the »Out of memory« message is displayed.