E-BAM-9805 Manual Rev B.docx
Page 25
3.5.6 Filter Sensors
This screen provides the options and indications needed to default, verify, and calibrate the filter
temperature, humidity, and upper and lower pressure sensors. See section 5.10 for detailed
instructions.
The Filter Sensor field indicates which sensor is
currently being audited. To select a different
sensor, press the green bordered value box and a
list of the sensors will be displayed. Use the arrow
buttons at the bottom of the screen to see more
choices. Tap the option to be used, and it will be
selected. Press the OK button to exit the screen
and audit the sensor.
Figure 3-18 The Filter Sensors Screen
The E-BAM field is the current value of the chosen Filter Sensor.
The Standard field is the field where the value shown on the traceable reference standard is
entered if the E-BAM value needs to be adjusted. To make a change, press the green bordered
value box and the numerical entry keypad (Figure 3-5) will be displayed. Enter the value currently
shown on the standard and press the OK key to set it. Press the CALIBRATE key to set any
changes. The E-BAM and Standard fields should now match.
The DEFAULT key can replace the current calibration factors with the factory default settings on
any of the four sensors. If difficulty is encountered during the audit process, press the DEFAULT
key to set that specific sensor back to factory values and then try the calibration again.
3.5.7 Span Mass Audit
This screen runs the Span Mass Audit. The sequence begins as soon the menu is entered. See
section 5.9 for detailed instructions.
3.5.8 Tape Test
This screen provides a means to verify proper
installation and operation of the filter tape and
motors. Press the grey TEST TAPE button to
advance the tape. The pass criteria is ≥ 12 mm.
The status of the tape, such as OK or if there is a
Tape Break error, will be displayed to the right of
the TEST TAPE button.
Figure 3-19 The Tape Test Screen
3.5.9 Inlet Heater
This screen allows manual operation of the inlet
heater assembly. Press ON to turn the heater on
and verify the element heats up as expected.
Press OFF to turn the heater off; verify it shuts
off and then cools down. Exiting this test screen
will also turn off the heater if it is left in the ON
state.
Figure 3-20 The Inlet Heater Test Screen