GT-521S Manual Rev D 

 

Page 8 

2.3. 

Default Settings 

The GT-521S comes with the user settings configured as follows. 

Parameter 

Value 

Sizes 

0.3 µm and 0.5 µm 

Sample Location 

Samples 

Single 

Sample Time 

60 seconds 

Sample Hold Time 

0 seconds 

Volume (concentration)  

CF (particles/ft

3

Count Mode 

Normal (cumulative) 

Temperature Units 

Baud Rate 

9600 

Serial Mode 

RS-232 

Alarm Limits  

Disabled 

2.4. 

Initial Operation 

Before operating the GT-521S for the first time, it is recommended that the unit be fully 
charged.  Information regarding charging the battery is found in section 6 of this manual.   

Complete the following steps to verify proper operation.   

1.  Slide the power switch up to turn on power.  
2.  Observe the Startup Screen for 2 seconds then the Operate Screen (Section 

4.2)

3.  Press Start / Stop key. The GT-521S will sample for 1 minute and stop. 
4.  Observe the counts on the display. 
5.  The unit is ready for use  

3. 

User Interface 

The GT-521S user interface is composed of a 7 button keypad and a LCD display.  The 
following table describes keypad functionality. Note, some keys have more than one 
function.  

Key 

Description 

 

 

Starts or stops a sample. 

 

Starts the unit printing when in the Print Data Screen. 

 

Recalls the selected data when in the Recall Data Screen. 

 

 

Loads the Menu Screen when not editing a field. 

 

Loads the Operate Screen when in the Menu Screen. 

 

Cancels editing a field and return the field to the original value.  

 

 

Loads the screen associated with menu item. 

 

Begins editing of the selected item. 

 

Stops editing a field and saves the changed value. 

 

 

Navigates up/down when not editing (Menu). 

 

Modifies field when editing. 

 

 

Navigates right / left 

 

Move through history buffer (Operate screen)

 

MENU

ESC

ENTER

STOP 

START 

Содержание GT-521S

Страница 1: ...GT 521S Manual Rev D Met One Instruments Inc Corporate Sales Service 1600 NW Washington Blvd Grants Pass OR 97526 Tel 541 471 7111 Fax 541 471 7116 www metone com service metone com GT 521S MANUAL...

Страница 2: ...tins are often posted on our website Please contact us and obtain a Return Authorization RA number before sending any equipment back to the factory This allows us to track and schedule service work an...

Страница 3: ...l Output 11 4 3 3 Samples 11 4 3 4 Sample Time 11 4 3 5 Hold Time 11 4 3 6 Count Mode 12 4 3 7 Sample Timing 12 5 Menu Selections 13 5 1 Edit Menu Items 13 5 2 Sample Setup Screen 13 5 2 1 Location Nu...

Страница 4: ...20 6 Charging the Battery 21 7 Serial Communications 22 7 1 Serial Interface 23 7 2 Commands 24 7 3 Real Time Output 24 7 4 Comma Separated Value CSV 25 7 5 Printer Format 25 8 Maintenance 26 8 1 Ser...

Страница 5: ...Unpacking Unpack and inspect the contents of the shipping container Standard items included are shown in Figure 1 Standard Accessories Optional accessories are shown in Figure 2 Optional Accessories...

Страница 6: ...rivers that are not compatible with this product To install USB drivers Insert the Comet CD The install program should run automatically and display the screen below If an AutoPlay pop up window appea...

Страница 7: ...PN GT 521S Battery Charger PN 390031 Power Cord PN 400113 Carrying Case PN 8517 Calibration Certificate PN GT 521S 9600 ISO Kinetic Probe PN G3110 Zero Count Filter PN G3111 USB Cable PN 500787 Comet...

Страница 8: ...k Input jack for the battery charger This connection charges the internal battery pack and provides continuous operating power for the unit USB Port For USB serial communication Serial Port For serial...

Страница 9: ...en Section 4 2 3 Press Start Stop key The GT 521S will sample for 1 minute and stop 4 Observe the counts on the display 5 The unit is ready for use 3 User Interface The GT 521S user interface is compo...

Страница 10: ...ent sample data and previous sample data Figure 5 shows the Operate Screen 18 AUG 06 11 23 0 3 0 CF 0 5 0 CF Location 001 Figure 5 Operate Screen w o Temp RH The top line of the Operate Screen shows d...

Страница 11: ...ount system to initialize HOLDING GT 521S is taking multiple samples and is waiting for the hold time to finish STOPPING GT 521S is stopping the sample and is waiting for the count system to stop COUN...

Страница 12: ...2 Real Time Serial Output GT 521S provides real time output on the serial port at the end of each sample The format of the output is controlled by the report type setting section 5 8 1 4 3 3 Samples T...

Страница 13: ...more detail in section 5 3 COUNTING 58 0 3 29 780 CF 0 5 1 400 CF Location 001 Figure 8 Differential Counts 4 3 7 Sample Timing The following figures depict the sample timing sequence for both single...

Страница 14: ...t the date and time SET CONTRAST Adjust the display contrast ABOUT Display firmware version and serial number 5 1 Edit Menu Items To change settings press Menu to display the main menu press or to nav...

Страница 15: ...seconds will increase pump life 5 2 4 Samples The samples setting controls the number of samples to take as illustrated below Selection Description REPEAT Repeat configures the unit for continuous sa...

Страница 16: ...er of particles between a given size and the size larger When in differential mode the volume units will be preceded the delta character on the screen For the serial output a lower case d character is...

Страница 17: ...ents to print Setting location to 000 prints all locations Settable from 0 999 01 JAN 00 The date time to begin printing sample events from 18 AUG 06 The date time to stop print samples at After the p...

Страница 18: ...ey while viewing the Memory Screen This will delete all the sample events in memory A warning screen will be displayed to prevent accidental erasure 5 7 Count Alarms Screen Configure count alarm limit...

Страница 19: ...annel 2 Hi Alarm 5 00 V 5 8 Serial Configure the serial port settings REPORT CSV BAUD 9600 SERIAL RS 232 Figure 20 Serial Screen 5 8 1 Report Type The Report setting determines the output format for t...

Страница 20: ...tting on the top of the unit PN 9801 2 Access the Set Flow Screen Figure 21 by pressing Menu then select Set Flow The pump will start automatically when you enter the Set Flow Screen and stop when you...

Страница 21: ...TIME 11 25 36 Figure 23 Set Clock Screen Date Time Formats Date dd mmm yy dd day mmm month yy year Time hh mm ss hh hours mm minutes ss seconds 5 12 Set Contrast Screen Press or to improve display qua...

Страница 22: ...ht side of the GT 521S The battery charger is universal and will work with power line voltages of 100 to 240 volts 50 to 60 Hz A discharged battery pack will take approximately 2 5 hours to fully char...

Страница 23: ...talled first Windows may install generic drivers that are not compatible with this product To install USB drivers Insert the Comet CD The install program should run automatically and display the scree...

Страница 24: ...with the GT 521S through the RS 232 DB 9 port requires a custom serial cable PN3228 Standard serial cables will not work and may cause damage to the instrument if connected Below is a description of t...

Страница 25: ...last n records If n is omitted this returns the last record D Date mm dd yy Set Date T Time hh MM ss Set Time C Clear data S Start a sample E End a sample ST Sample Time RV Show software revision ID S...

Страница 26: ...nnel 2 Size 00 5 5 Channel 2 Count TC L CF M3 00008680 Note 2 6 Temperature C F 022 Note 2 Note 3 7 RH 033 Note 3 8 Location 001 9 Sample Time 0 9999 seconds 0060 10 Status Bits see below 000 Note 5 1...

Страница 27: ...igure the unit as follows Samples SINGLE Sample Time 60 seconds Volume Total Count TC 3 Start and complete a sample 4 The smallest particle size should have a count 1 8 3 Flow Rate Test The flow rate...

Страница 28: ...to laser radiation which can cause eye injury Symptom Possible Cause Solution Display does not turn on Low Battery Defective Battery Charge battery Send to service center Pump does not turn on when a...

Страница 29: ...ery Life 24 hours Intermittent operation 6 hours continuous operation Battery Charge Time Fully changed in 2 5 hours AC Adapter Charger Li ion battery charger 100 240 VAC 50 60Hz Communications USB RS...

Страница 30: ...GT 521S Manual Rev D Page 29 11 Warranty Service...

Отзывы: