![MessKonzept FTC320 Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/messkonzept/ftc320/ftc320_operating-manual_1775514012.webp)
Page 11 of 43
Mea-
suring
Gas
Carrier
Gas
Basic
range
Smallest range
Smallest
supressed zero
range
Multi Gas
Mode
H
2
N
2
/ air
0% - 100%
0% - 0.5%
98% - 100%
Yes
H
2
Ar
0% - 100%
0% - 0.4%
99% - 100%
Yes
H
2
He
20% - 100%
20% - 40%
85% - 100%
On
request
H
2
CH
4
0% - 100%
0% - 0.5%
98% - 100%
On
request
H
2
CO
2
0% - 100%
0% - 0.5%
98% - 100%
On
request
He
N
2
/ air
0% - 100%
0% - 0.8%
97% - 100%
Yes
He
Ar
0% - 100%
0% - 0.5%
98% - 100%
Yes
CO
2
N
2
/ air
0% - 100%
0% - 3%
96% - 100%
Yes
CO
2
Ar
0% - 60%
0% - 10%
-
Yes
Ar
N
2
/ air
0% - 100%
0% - 3%
96% - 100%
Yes
Ar
CO
2
40% - 100%
-
80% - 100%
Yes
CH
4
N
2
/ air
0% - 100%
0% - 2%
96% - 100%
Yes
CH
4
Ar
0% - 100%
0% - 1.5%
97% - 100%
Yes
O
2
N
2
0% - 100%
0% - 15%
85% - 100%
Yes
O
2
Ar
0% - 100%
0% - 2%
97% - 100%
Yes
N
2
Ar
0% - 100%
0% - 3%
97% - 100%
Yes
N
2
CO
2
0% - 100%
0% - 4%
96% - 100%
On
request
NH
3
H
2
0% - 100%
0% - 5%
95% - 100%
On
request
CO
2
H
2
0% - 100%
0% - 2%
99% - 100%
On
request
SF
6
N
2
/ air
0% - 100%
0% - 2%
96% - 100%
On
request
Table 2.1: Measuring ranges of typical gas compositions for analysis with the FTC320.
The FTC320 must not be used with explosive gases. Flammable gases such as H
2
and
CH
4
may only be used in devices filled with glass balls. A gas mixture of a flammable
gas with an inert gas in a mixing ratio such, that it is still inflammable for any amount of
air added is called totally inertised. Totally inertised gases can also be used in devices
without glass balls.
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V02.pdf
Содержание FTC320
Страница 45: ......