8.4.2
Analog Output 1
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Current Loop
Analog Out 1
Analog Out 2
Adjust Loop
AOut1-Param
AOut1-Offset
AOut1-Gain
AOut1-Param
set: 0.00000
new: 0.00000
ESC
/OK
AOut1-Param
AOut1-Offset
AOut1-Gain
AOut1-Param
AOut1-Offset
AOut1-Gain
AOut1-Offset
set: 0.00000
new: 0.00000
ESC
/OK
AOut1-Gain
set: 0.00000
new: 0.00000
ESC
/OK
Figure 8.16: Setup of Analog Output 1
“Analog Out 1“ is a non DC-isolated 0 to 10V output. The maximum load is 1kOhm.
Several parameters can be routed to the analog output. In the “AOut-Parameter“-menu you can
select the signal you wish to route to the analog output by entering the value number. If you wish
to route the linearized measuring signal, select 1.00000. Please contact Messkonzept for a list of
paramters available in your device.
“AOut1-Offset“ and “AOut-Gain“ allow scaling of the analog output according to a linear equation:
Output = AOut1Gain
·
AOut1Offset.
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf