Page 32 of 58
8.2.4
Multi Gas Mode List (only with MGM-option)
A1 A2 Er
0.30
[%]
H2/N2
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Operation
Calibration
Diagnosis
Setup
Unit -> Vol%
H2 in N2
T90 Response
List for MGM
H2 in N2
------------
Not in list
ESC/OK
H2 in N2
------------
Not in list
ESC/OK
O2 in N2
------------
Not in list
ESC/OK
User poly.
------------
Not in list
ESC/OK
H2 in N2
------------
Is in list
ESC/OK
Figure 8.5: gas pair selection in Multi Gas Mode
Devices supporting the Multi Gas Mode (MGM) can measure (in succession) up to 16 different bi-
nary gas mixtures following your wishes when ordering the FTC300. From the list of all available
binary gas mixtures a list for the short-key access from the operational screen, further called “List for
MGM“, can be edited. Binary gas mixtures can be added to or removed from the “List for MGM“; gas
mixtures labelled with “Is in list“ are included in the short list. Toggle through the available binary gas
mixtures using the <ENTER> or the <UP> key. The <RIGHT> key changes to the third display line,
where the <ENTER> key is used to switch between “Is in list“ and “Not in List“. Press <RIGHT> to
select “OK“ and <ENTER> to confirm the selection.
From the operational screen (main screen showing the currently measured gas concen-
tration) you can quickly toggle through the set-up MGM short list: Press <RIGHT> once
to activate the quick access and then press <ENTER> until you read the desired gas
pair name from the short list in the bottom right corner of the screen. Pressing <RIGHT>
again deactivates the quick access.
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf
Содержание FTC300
Страница 1: ...FTC300 Gas analysis using thermal conductivity measurement Operating Manual ...
Страница 55: ...Page 54 of 58 This page is intentionally left blank File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...
Страница 58: ...Page 57 of 58 Chapter 12 Appendix Dimensional Drawing File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...