Weight Procedure
The Weight Procedure is used to calibrate the scale to show a correct net weight.
To run this procedure, you will need calibration weights that represent at least 20% of the
Design Weight
. The weight in the hopper plus the sum of the calibration weights may not
exceed the
Hi Weight
limit. The procedure will first sample the current weight, and then
the weight with the calibration weights added.
To begin the Weight Procedure push the
Start
button.
If the total weight (weight in the hopper plus calibration weights) exceeds the
Hi Weight
limit, the display will show an Overload Condition screen.
Normally, the start weight is sampled. This requires that the weight value is stable.
In the next screen, the sampled weight before applying the calibration weighs is
displayed (
Test Zero
), along with the expected calibration weight. By pushing the Change
Value button, you can alter the Calibration Weight value.
Push the OK button and load the calibration weights onto the hopper.
35.00.EX O&M Manual
49