 Perfusion Solution for Process Development User Guide 

21

Symptom

Corrective Action

System leak

Check for cuts, cracks, or damage to assembly
Ensure system pressure is less than 5 psi
Ensure flowmeter clamp does not pinch the 

tubing

Unexpected or 

unstable pressure 

readings

Ensure cable connections are good
Proper installation of P1 on Feed, P2 on 

retentate, and P3 on permeate
Perform zero adjustment prior to run, after 

priming
Clear air from the sensor

Unexpected 

or unstable 

flowmeter 

readings

Ensure cable connections are good
Clean flowmeter and tubing with IPA
Close clamp around tubing and ensure 

flowmeter clamp does not pinch the tubing 
 Ensure pump is running and flow is moving
Prime tubing in flowmeter, remove air bubbles
Choose correct calibration set in system settings 

(tubing type, temperature) 
 Perform zero adjustment at the correct 

temperature, prior to run and after priming
Re-zero sensor
Check flowmeter calibration against reference 

flowmeter

Pump unable to 

achieve target 

flow rate

Re-prime tubing to remove air bubbles
Remove pump head from pump to clear air from 

inside
Check for tubing pinch or kink
Check for filter fouling (high P1 with low P2)
Check dip-tube placement and ensure it does 

not pull air from sparger
Check system mode, “flow control” vs “speed 

control”

No flow through 

permeate

Ensure permeate line is not clamped or pinched

Symptom

Corrective Action

Early filter fouling 

(High P1 or low 

P3)

Check that crossflow rate is at setpoint
Check system mode, “flow control” vs “speed 

control”
Ensure maximum flux is within recommended 

range
Swap out filter
Check flowmeter calibration setting is correct
Check flowmeter calibration against reference 

flowmeter

System pressure 

increase (High P1 

and P2)

Check bioreactor pressure (ex. vent filter 

clogging)
Check for tubing pinch or kink on retentate line
Check flowmeter calibration setting is correct
Check flowmeter calibration against reference 

flowmeter

Cell growth issues 

in bioreactor

Check flow rate and pump RPM are in 

recommended range
Check flowmeter calibration setting is correct
Check for cell loss in permeate line
Prime filter during media hold only
Re-prime system to remove air bubbles
Check flowmeter calibration against reference 

flowmeter

Cells found in 

permeate

Swap out filter

Unexpected or 

undesired alerts

Adjust alert settings
Adjust alarm limits
Perform factory reset
Follow recommendations provided for each alert

Data logging is 

not saved

Ensure USB is inserted during run
Ensure USB drive has adequate space
USB is not removed until data logging is turned 

off and software indicates it is safe to remove

Содержание Milipore Cellicon Perfusion Solution for Process Development

Страница 1: ...n Solution for Process Development User Guide Contents Introduction 2 System Components 2 Setting Up the System 3 Starting the System 4 User Interface 4 Settings 7 Running the Pump 7 Alerts 9 Settings...

Страница 2: ...scribes the installation setup and use of the CelliconTM Perfusion Solution for Process Development This system is for industrial use only and cannot be used in a residential environment System Compon...

Страница 3: ...ors on the assembly as follows Press 1 to the feed line Press 2 to the retentate line Press 3 to the perfusate permeate line Connecting the Bioreactor Use a tubing welder to connect The cell retention...

Страница 4: ...on Powering up the system launches the perfusion control software The initial screen is the username selection screen After selecting the username the password screen will appear Touch the white box o...

Страница 5: ...ton Selecting the Settings button opens the settings screen User Icon and Time Date Display Selecting the user icon allows the user to lock the screen or switch users Lock screen will cause the contro...

Страница 6: ...low Setpoint Prime System Zero Adjust Flow Rate Speed Setpoint Pump Start Decrease Flow Rate Increase Flow Rate Status The current values state of the system components and data logging switch are dis...

Страница 7: ...ton will read START PUMP If the pump is on the button will read STOP PUMP Stopping the pump will cause a confimation screen to appear Prime System The pump needs to be primed to reduce damage to cells...

Страница 8: ...t be stopped A pop up will appear to confirm that the pump is stopped Flowmeter 2 does not require that the pump is stopped All sensors should be zero adjusted prior to starting a run Only one sensor...

Страница 9: ...splayed Alarm Acknowledgement Alerts can be removed from the main screen by selecting Acknowledge from the alert bar or in the alert log The alert will reappear on the alert bar if it remains active 1...

Страница 10: ...ill make a loud beep when the alert is activated Select Dismiss to return to the main screen then select Acknowledge on the alert bar or within the alert log The data region can still be viewed when a...

Страница 11: ...Critical Alarm High Optional 1 Pressure Alarms Pressure Critical Alarm Low Optional 1 Pressure Warning Alarm Low Never 2 Pressure Warning Alarm High Never 2 Pressure Critical Alarm High Optional 1 Al...

Страница 12: ...ending is fixed and generated with the date followed by the time when the logfile is created When data logging is switched On a prompt to enter a filename will appear This filename will replace the t...

Страница 13: ...error The resolution of the flow signal can be improved by decreasing the full scale of the flowmeter The resolution of the flow signal is always 1 10 000 of the full scale Decreasing the full scale b...

Страница 14: ...Refer to Table of Alerts for more information Critical alarms are shown in red color type 1 alert Warning alarms are shown in yellow color type 2 alert Neither critical nor warning alarms will shut d...

Страница 15: ...nalog Input PLC Digital Input PLC Analog Output User Accounts Settings The administrator can generate delete and edit user accounts including password protection capabilities and rescrict access permi...

Страница 16: ...s have been added to the reactor This will eliminate the possible introduction of cell shear when the system is primed see Priming Preparation Equipment Needs The following user supplied items may be...

Страница 17: ...r or gas in the pump head will negatively affect pump performance and decrease the recirculation flow rate MobiusTM 3L Single use Bioreactor When using the MobiusTM single use 3L reactor the inlet to...

Страница 18: ...lect Prime system on the main menu screen This will switch the controller into Speed setpoint mode and provide step by step guidance on how to complete the priming 2 Ensure that the perfusate permeate...

Страница 19: ...ved on the USB drive the log interval and the log file size Choose a log interval of 60 seconds or greater and a log file size of 65000 lines Insert the USB drive into a computer to view data Open Mic...

Страница 20: ...ars light grey and does not read values Connect cables from component to controller Install correct software Component is in an alarm state appears red or yellow Plug in component cables Pump alert Pl...

Страница 21: ...d ensure it does not pull air from sparger Check system mode flow control vs speed control No flow through permeate Ensure permeate line is not clamped or pinched Symptom Corrective Action Early filte...

Страница 22: ...un Disposing of the Filter The filter is completely single use and should be disposed of after use Follow company procedure for waste disposal Cleaning the Hardware All parts of the system should be w...

Страница 23: ...s All other trademarks are the property of their respective owners Detailed information on trademarks is available via publicly accessible resources MK_UG5285EN Rev 1 06 2020 2020 Merck KGaA Darmstadt...

Отзывы: