6.4.3.7 Rate Stability Parameters
Rate parameters for the control rate can be found in the Control Settings App and can be configured us-
ing the Rate Stable Window and Rate Stable Delay buttons. The Rate Stable Window button allows the
user to enter a value as a percentage of the active range of transducer. This value represents the pres-
sure window within which the control rate value would be considered stable by the user. The Rate Stable
Delay button lets the user add a desired delay until the control rate is considered stable while being in the
rate stable window.
Rate
Parameters
Figure 6.4.3.7 – Rate Parameters