6.4.9 Remote Application
With the Remote Settings application user can select the remote command set for all interfaces.
The GPIB address, Ethernet network parameters and Serial parameters can also be set here.
Details about the Remote Operation (command sets, cable requirements, etc.) can be found in
Section 7, Remote Operation.
Figure 6.4.9 – Remote Application
6.4.9.1 Remote Command Set
The remote command set parameter provides a choice of the Mensor command set or the WIKA SCPI /
DPI command set (Figure 6.4.9.1 -A). Both sets of commands are listed in Section 7, Remote Operation.
Figure 6.4.9.1 - A – Remote Command Set
The DPI5xx command also provides an option to set the Checksum (Figure 6.4.9.1 - B) which when set
to ON or AUTO appends a mod 100 checksum to end of any remote communication response. A proper
mod 100 checksum must be sent to the instrument when the checksum is ON. Sending the checksum is
optional when set to AUTO. The flow control is set to NONE as default internally. The termination char
-
acter sets the output termination character for all remote communication command sets (Figure 6.4.9.1
- C). The commands are listed in Section 7, Remote Operation.