background image

page 2

General notes, safety notes, notes on electrical equipment

and electromagnetic compatibility, as well as

notes on maintenance of the unit are to be found in the

Operating Instructions for the waterbath.

Brief technical description

The shaking device permits controlled shaking of beakers or test tubes in a waterbath.
The shaking device performs a reversing linear movement with a frequency of approx. 35 to 160 strokes 
per minute. The shaking speed is controlled electronically and can be adjusted in 10 speed steps. The stroke 
is approx. 15 mm.
The ambient conditions for its operation are the same as for the basic units.

Inserting and removing

The bath can be operated without shaking device if required. After the shaking device and the support 
basket has been removed, the side panel supplied should be secured to the gable cover by means of the 
two screws in orded to ensure optimum sealing of the cover. Fitting the shaking device has to take place 
in the reverse order.

Safety note
If the shaking device has to remain in a waterbath that is switched on, it has to remain switched on as 
well! The shaking speed can be set to a standstill! When being switched off, the overpressure steam 
blockage is out of action and condensate may pass into the housing of the shaking device!