3 Setup and installation
13
Installation examples
The following pages provide examples for the recommended installation types for the feed water supply.
The connection of a different water treatment unit producing the same water quality is possible after con-
sultation with MELAG.
Example 1 – feed water supply from the internal storage tank (delivery
state)
Direct feed water supply from the internal storage tank without an additional water connection.
No further installation is required apart from the electrical connection.
The wastewater is collected in the left-hand side chamber of the storage tank and is to be emptied
manually.
A float switch installed in the storage and wastewater tank notifies insufficient feed water or a full wastewa-
ter chamber.
4
3
2
1
5
Pos. Description
Art. no.
1
Silicon hose (9/6 mm, 5 m)*
)
ME34410
2
Plug for water drain*
)
ME31140
3
Feed water inlet hose*
)
ME34410
4
Feed water connection, Vacuklav*
)
ME21175
5
Mains connection**
)
--
*
)
present on the device side
**
)
present on the building side