Chapter 8 – Malfunctions 

18 

Chapter 8 – Malfunctions 

 

 

WARNING 

Incorrect opening can compromise electrical safety and pose a danger to the user. 

 

To avoid the risk of electrical shocks, never open the housing of the sealing device. 

 

The device may only be serviced by a technician authorized by the MELAG technical support 
department. 

 

 

PLEASE NOTE 

If an error message is displayed during a sealing procedure, this must be cleared by pressing the

 key. Normal operation can then be continued. 

 

 

Incident 

Possible cause 

Remedy 

No display on the 
screen fan is not 
functioning. 

The sealing device is not switched 
on at the power switch. The power 
cable is not connected or is not 
connected with the mains socket. 

Switch on the sealing device at the mains 
switch (see p.6 Fig. 1/(7)), check the 
connection of the power cable from the sealing 
device to the mains socket. On repeated 
occurrence, inform your MELAG customer 
services. 

The required sealing 
temperature is not 
reached. 

The seal seam is 
burned. 

The sealing temperature has not 
been  
set correctly. 

Check the sealing temperatures set on the 
sealing device and if necessary, change the 
setting (see page 12). On repeated 
occurrence, inform your MELAG customer 
services. 

The seal seam is not 
correct. 

The sealing device is not capable 
of generating the sealing 
temperature recommended for the 
packaging. 

Check the sealing temperatures set on the 
sealing device and if necessary, change the 
setting (see page 12). 
On repeated occurrence, inform your MELAG 
customer services. 

Seal seam is irregular  
and contains defects. 

The packaging material is not 
suitable for use with this sealing 
device. The pressure roll is soiled. 
The pressure roller is soiled. 

Use packaging material suitable for this sealing 
device (see seal seam strength test). The 
pressure roller requires cleaning by the 
MELAG customer services. 

The conveyor belt 
does  
not start. 

The sealing device has not yet 
reached operating temperature. 

Wait until the sealing device has reached 
operating temperature. 
On repeated occurrence, inform your MELAG 
customer services. 

The display flashes. 

The energy supply has been 
interrupted. 

Please consult your MELAG customer 
services. 

 

 

Содержание MELAseal Pro

Страница 1: ...reatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing competence in hygiene and Quality made in Germany we guarantee that the...

Страница 2: ...a component of the product About this manual Symbols used Symbol Explanation Indicates a dangerous situation which if not avoided could entail slight to life threatening injuries Indicates a dangerou...

Страница 3: ...t medium 13 MELAflash CF Card Printer as output medium 13 Using the computer as an output medium 13 Opening and printing log files on the computer 14 Log folder structure 14 File name and ending 14 Re...

Страница 4: ...other than a member of a MELAG authorized technical customer service Fan The fan in the base of the sealing device must be free at all times and may not be allowed to become blocked or congested Failu...

Страница 5: ...al Device Directive Suitable materials It is suitable for the heat sealing of the following packaging materials Heat sealable transparent sterilization packaging e g MELAfol Should you wish to use pac...

Страница 6: ...seal Pro rotary sealer Power cable Operating Manual Views of the device Fig 1 Views of the device 1 Guide rail for straight introduction of the sterilization packaging 2 Operating and display field 3...

Страница 7: ...increase or reduce the sealing temperature Reverse key for removing the sterilization packaging upon a blockage of the conveyor belts Daily batch counter for increasing or reducing the total number o...

Страница 8: ...hould be located at least 1 5 m away radius from the treatment area Observe all the information contained in this chapter during commissioning DANGER Failure to comply with the set up conditions can r...

Страница 9: ...aling device 1 Loosen the screw a little in order to slide the guide rail forwards or backwards The scale to the right of the adjusting screw displays the overhang in mm measured from the outer edge o...

Страница 10: ...is could result in the deposit of film residue on the sealing rail Using the reverse key Should either the sterilization packaging become jammed in the device or the conveyor belt come to a halt you c...

Страница 11: ...the current date proceed as follows 1 Activate the sealing device at the power switch 2 Press as soon as the display shows the current software version 3 Press repeatedly to access the Date menu 4 Pr...

Страница 12: ...the Signal tone It is possible to set a signal tone to control the number of packages to be sealed It will sound after the pre set number of seals has been completed This function is useful when need...

Страница 13: ...ELAflash CF Card Printer from software version 1 9 supports the recording of logs for the MELAseal Pro rotary sealer If you want to employ the MELAflash CF Card printer as the output medium connect it...

Страница 14: ...Card and the computer The directory name consists of five characters identical with the first five characters of every log e g 9K0EQ Under this folder is a sub directory containing the month of the pr...

Страница 15: ...ace Example of a log row for a successfully concluded sealing procedure 01161 2009SPF1000 09 12 15 14 28 180 099 88 185 175 120 090 120 078 F300 Key The following parameters are displayed successively...

Страница 16: ...actical and inexpensive solution suitable for sealing light packaging Roll dispenser The roll dispenser represents the ideal assistant when you need to process MELAfol transparent sterilization packag...

Страница 17: ...t suitable for use with medical products It should not deposit any oily residue After 1 year Clean the fan on the base of the sealing device once a year in order to prevent it from becoming clogged wi...

Страница 18: ...rned The sealing temperature has not been set correctly Check the sealing temperatures set on the sealing device and if necessary change the setting see page 12 On repeated occurrence inform your MELA...

Страница 19: ...customer services Temp too high The sealing temperature is too high for a short period If the malfunction message disappears after a few minutes you can continue operation as usual On repeated occurr...

Страница 20: ...ion documentation within the scope of process clearance Check for dryness and integrity once again before using the instruments Weekly Checking the seal seam with aids Perform a visual control using M...

Страница 21: ...st price of 125 00 plus VAT state 01 2017 After the film test strips have been tested MELAG will issue verification of testing and if successful a certificate This certificate confirms conformity of t...

Страница 22: ...omposite film should be filled with a max of up to 3 4 of its volume DIN 58953 7 2010 When using transparent sterilization packaging from a roll the removal side must have an overlap of min 1 cm betwe...

Страница 23: ...thus not be generally specified Table 2 only makes recommendations regarding the storage length of sterile medical products The following requirements apply to the storage of sterile medical products...

Страница 24: ...dapter 80270 Replacement parts Separating plate with roll dispenser 71200 Angle adapter for RS232 socket 80210 Device fuses 3 15 A T 5x20 74320 Technical Data Model name MELAseal Pro Device dimensions...

Страница 25: ......

Страница 26: ...a e 6 10 10829 Berlin Germany E Mail info melag de Web www melag com Responsible for content Technical Office We reserve the right to technical alterations Your stockist BA_sealPro_EN_v6 docx Rev 6 17...

Отзывы: